Background And Purpose: Polyethylene wear is an important cause of aseptic loosening in hip arthroplasty. Detection of significant wear usually happens late on, since available diagnostic techniques are either not sensitive enough or too complicated and expensive for routine use. This study evaluates a new approach for measurement of linear wear of metal-backed acetabular cups using CT as the intended clinically feasible method.
Material And Methods: 8 retrieved uncemented metal-backed acetabular cups were scanned twice ex vivo using CT. The linear penetration depth of the femoral head into the cup was measured in the CT volumes using dedicated software. Landmark points were placed on the CT images of cup and head, and also on a reference plane in order to calculate the wear vector magnitude and angle to one of the axes. A coordinate-measuring machine was used to test the accuracy of the proposed CT method. For this purpose, the head diameters were also measured by both methods.
Results: Accuracy of the CT method for linear wear measurements was 0.6 mm and wear vector angle was 27 degrees . No systematic difference was found between CT scans.
Interpretation: This study on explanted acetabular cups shows that CT is capable of reliable measurement of linear wear in acetabular cups at a clinically relevant level of accuracy. It was also possible to use the method for assessment of direction of wear.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1080/17453670710015003 | DOI Listing |
Hip Int
January 2025
Department of Orthopaedic Surgery, The Ottawa Hospital, Ottawa, Ontario, Canada.
Background: Different methods can help to optimise sagittal cup orientation in total hip arthroplasty (THA) based on individual spinopelvic characteristics. This study aimed to: (1) assess how often combined sagittal index (CSI) and hip-spine-classification targets were achieved post THA; (2) compare anteversion/inclination between cups in-/outside optimal CSI zone; and (3) determine association with outcome.
Methods: This is a multicentre, prospective, case-cohort study of 435 primary THA for osteoarthritis (53% females; age: 65 ± 12 years; follow-up: 2.
Hip Int
January 2025
Orthopaedic Clinic Lucerne, Lucerne, Switzerland.
Introduction: The RM Pressfit vitamys is an uncemented, titanium particle-coated, isoelastic monoblock cup made of vitamin E blended highly cross-linked polyethylene. We addressed the following questions: (1) What are the clinical and (2) radiographic outcomes 10 years after implantation? (3) What is the revision rate?
Methods: In this prospective observational study in a tertiary care centre we investigated all consecutive cases of total hip replacement with the RM Pressfit vitamys cup between September 2009 and November 2011. It was implanted in 162 hips, 49.
Arch Orthop Trauma Surg
December 2024
Virtua Health, Marlton, NJ, USA.
Objective: The primary research aim was to determine if the use of traditional or 3D printed prosthesis resulted in better functional outcome scores in hip arthroplasty.
Methods: A systematic review and meta-analysis was conducted utilizing the PRISMA 2020 guidelines. Six databases (PubMed, Embase, Scopus, WebOfScience, and Cochrane Library, Google Scholar) were searched yielding 1117 article titles and abstracts.
SICOT J
December 2024
The Hazeley Academy, Emperor Drive, Hazeley, Milton Keynes, MK8 0PT, United Kingdom.
Introduction: Robot-assisted instrumentation during total hip arthroplasty (THA) has the potential to improve acetabular cup positioning. This study aimed to evaluate the precision of robotic-arm-assisted THA (rTHA) and assess whether the system can provide accurate cup positioning comparable to conventional THA (cTHA).
Methods: A single-center prospective cohort study consisting of 151 patients who underwent THA (108 rTHA and 43 cTHA).
Med Biol Eng Comput
December 2024
Sch Mech Engn, Southwest Jiaotong Univ, Chengdu, 610031, People's Republic of China.
The demand for total hip replacement surgery is increasing year by year. However, the issue of hip prosthesis failure, particularly the modular acetabular cup, still exists. The performance and functional requirements of modular acetabular cups have not yet met clinical expectations.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!