A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Biomechanical and histologic evaluation of the Norian craniofacial repair system and Norian Craniofacial Repair System Fast Set Putty in the long-term reconstruction of full-thickness skull defects in a sheep model. | LitMetric

Biomechanical and histologic evaluation of the Norian craniofacial repair system and Norian Craniofacial Repair System Fast Set Putty in the long-term reconstruction of full-thickness skull defects in a sheep model.

Plast Reconstr Surg

Cleveland and Akron, Ohio From the Departments of Plastic Surgery and Pathology and Plastic Surgery Research, Cleveland Clinic, and the Department of Biomedical Engineering, University of Akron.

Published: May 2008

Background: The purpose of this study was to investigate the long-term (12 months) strength and osteoconductive properties of two forms of carbonated calcium phosphate cements (i.e., the Norian Craniofacial Repair System and Norian Craniofacial Repair System Fast Set Putty) and to compare these two bone cement forms to an autogenous cranial bone graft in a full-thickness skull defect adult sheep model.

Method: Twenty-six sheep were assigned to one of eight groups (n = 3 per group). A 4.5-cm2 full-thickness defect was created in the right and left parietal bones. Reconstruction was performed with a full-thickness cranial bone autograft, the Craniofacial Repair System, or Fast Set Putty. Skull samples were harvested at day 1, 6 months, and 12 months. Biomechanical testing was performed using a vertical drop test.

Results: Although the Craniofacial Repair System and Fast Set Putty osseointegrated fully, there was little osteoconduction at 12 months. The Craniofacial Repair System was the weakest reconstruction, presenting the lowest peak force transmission and the highest displacement at 12 months. Fast Set Putty showed significantly higher values for peak force transmission and lower displacement when compared with the Craniofacial Repair System.

Conclusions: The Craniofacial Repair System progressively lost strength over the course of this study. Fast Set Putty demonstrated greater strength and rigidity than the Craniofacial Repair System. Both implants had limited bone ingrowth from defect borders, but both cements osteointegrated completely. Bone grafts regained biostructural characteristics and strength similar to those of intact bone and clinically performed the same in this sheep model.

Download full-text PDF

Source
http://dx.doi.org/10.1097/PRS.0b013e31816a9fd1DOI Listing

Publication Analysis

Top Keywords

craniofacial repair
40
repair system
36
fast set
24
set putty
24
norian craniofacial
16
system fast
16
craniofacial
10
repair
10
system
9
system norian
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!