Background: The PAS-Port (Cardica Inc, Redwood City, CA) is an automated system that allows for the clampless anastomosis of vein grafts to the aorta. The intent of this study was to prospectively compare one year graft patency of this system with conventional hand-sewn anastomoses in a prospective, randomized trial.

Methods: A total of 99 patients undergoing elective off-pump coronary bypass surgery were randomized to receive their proximal anastomoses with either the hand-sewn conventional technique or with the PAS-Port system. Patient follow-up consisted of multislice computed tomographic scans performed at discharge and one year postoperatively.

Results: Three patients had to be converted to on-pump due to technical reasons. Fifty-one patients were randomly assigned to the PAS-Port group and 48 patients to the control group. In five patients in the control group severe atherosclerosis of the aorta required cross-over to the use of the PAS-Port device, and in one patient in the PAS-Port group conversion to a hand-sewn anastomosis. Sequential anastomoses were performed in 88% of the control group and 73% of PAS-Port group grafts. Time needed for completion of the proximal anastomosis including graft loading was 187 +/- 19 seconds for the PAS-Port group and 406 +/- 34 seconds for the control group (p < 0.001). One patient died unrelated to cardiac events due to septic multiorgan failure and one stroke was observed in the control group. There was a trend toward a lower rate of postoperative delirium in the PAS-Port group (11.7% vs 25%, p = 0.088). Patency at discharge (100% PAS-Port group vs 97.8% control group) and after one year (97.8% PAS-Port group vs 93.7% control group) were comparable. One patient of the PAS-Port group died during long-term follow-up as a result of a severe stroke due to cerebrovascular disease.

Conclusions: This prospective randomized study demonstrated excellent short and midterm patency in both the hand-sewn and PAS-Port grafts. The PAS-Port system allowed for the rapid, safe, and effective creation of a proximal anastomosis without the need to clamp the aorta. Based on this study we consider this product a valid alternative for proximal anastomosis, especially in patients with severe aortic disease, to avoid side clamping of the aorta.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.athoracsur.2008.01.074DOI Listing

Publication Analysis

Top Keywords

pas-port group
32
control group
28
group
15
pas-port
14
prospective randomized
12
proximal anastomosis
12
pas-port system
8
group patients
8
patients control
8
patient pas-port
8

Similar Publications

Study Objective: To evaluate the long-term outcomes of clampless off-pump coronary artery bypass grafting (C-OPCAB) compared with conventional on-pump double clamping coronary artery bypass grafting (C-CABG).

Methods: From October 2006 to December 2011, 366 patients underwent isolated coronary artery bypass grafting. After propensity score matching of preoperative variables, 143 pairs were selected who received C-OPCAB with the use of device-assisted PAS-Port proximal venous graft anastomoses or C-CABG, performed by the same surgeon experienced in both techniques.

View Article and Find Full Text PDF

Unlabelled: Backgrounds:The purpose of this study was to compare the long term patency of PAS-Port system with other hand-sewn anastomosis system.

Methods: From 2009 to 2018, a total of 79 anastomoses were done in coronary bypass surgery with PAS-Port system, and a total of 252 anastomoses with Enclose Ⅱ. Among them, 76 anastomoses with PAS-Port system (group P) and 246 anastomoses of enclose Ⅱ (group E) were evaluated using angiography or multislice computer tomography.

View Article and Find Full Text PDF

Objectives: Meta-analyses from observational and randomized studies have demonstrated benefits of off-pump surgery for hard and surrogate endpoints. In some of them, increased re-revascularization was noted in the off-pump groups, which could impact their long-term survival. Therefore, we analyzed the course of all patients undergoing isolated coronary surgery regarding the major cardiac and cerebrovascular event (MACCE) criteria.

View Article and Find Full Text PDF

Anaortic coronary artery bypass proved to prevent early neurologic injury compared to on-pump CABG. The Cardica PAS-Port(®) is a fully automated device that might be able to perform proximal aorto-venous anastomoses without an increased embolic risk. We evaluated early post-operative neurologic outcome in a matched population following clampless OPCAB (CCAB: either "all-arterial" or with automatically anastomosed venous grafts) or on-pump CABG.

View Article and Find Full Text PDF

Purpose: We examined grafts employing for morphological analysis of early and long-term results on proximal anastomosis with the PAS-Port(®).

Methods: One hundred and four patients treated by OPCAB with PAS-Port(®) were performed postoperative MDCT. Morphological evaluation of the proximal anastomotic region was classified into three groups (A; graft was anastomosed almost perpendicularly to the aortic wall, B; graft was same type A, but subsequently curved to form an acute angle with the aortic wall, C; graft take off acute angle with the aortic wall) evaluated on planar and sagittal sections.

View Article and Find Full Text PDF