Background And Purpose: Intracranial aneurysm (IA) treatment through hemodynamic modification with novel stent designs is a burgeoning area of research. We present a feasibility study for a new low-porosity patch-containing stent designed to treat intracranial aneurysms. The device is deployed so the patch covers the aneurysm neck ensuring strong flow diversion away from the aneurysm while keeping a low probability of occlusion of perforating vessels.

Methods: We created 17 side-wall aneurysms in 6 dogs, 2 per carotid artery if animal size permitted. Twelve proximal aneurysms were treated with AVSs: 5 distal aneurysms were untreated, serving as controls against self-thrombosis; 7 treated aneurysms were fully-covered; and 5 were partially-covered. After 4 weeks, a final angiogram was performed and aneurysms were explanted. Angiograms acquired pre- and posttreatment and at 4-week follow-up were analyzed quantitatively using normalized time-density curves (NTDC). Cone-beam micro-CT and histological specimen analysis were then performed.

Results: Posttreatment, NTDC average peaks dropped to 45% of initial values for the partially-covered aneurysms and 78% for the fully-covered aneurysms. Cone-beam micro-CT imaging performed at 4 weeks posttreatment showed partial thrombosis in 4 of 5 partially-covered aneurysms and complete thrombosis in all fully-covered aneurysms. Histology revealed neointimal coverage of all asymmetrical patch regions and thrombus formation in both fully- and partially-covered aneurysms. Four-week follow-up was not done for 1 animal (2 controls, 2 treated) that expired because of groin hemorrhage and for another animal (1 aneurysm) with an occluded carotid.

Conclusions: We demonstrate aneurysmal blood flow diversion using a new low-porosity patch-containing asymmetrical vascular stent in a canine side-wall aneurysm model. Overall results are encouraging and support continued AVS development.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2562235PMC
http://dx.doi.org/10.1161/STROKEAHA.107.503862DOI Listing

Publication Analysis

Top Keywords

low-porosity patch-containing
12
partially-covered aneurysms
12
aneurysms
11
vascular stent
8
feasibility study
8
study low-porosity
8
patch-containing stent
8
flow diversion
8
cone-beam micro-ct
8
fully-covered aneurysms
8

Similar Publications

Asymmetric vascular stent: feasibility study of a new low-porosity patch-containing stent.

Stroke

July 2008

Department of Neurosurgery, School of Medicine and Biomedical Sciences, University at Buffalo, State University of New York, Buffalo, NY 14209, USA.

Background And Purpose: Intracranial aneurysm (IA) treatment through hemodynamic modification with novel stent designs is a burgeoning area of research. We present a feasibility study for a new low-porosity patch-containing stent designed to treat intracranial aneurysms. The device is deployed so the patch covers the aneurysm neck ensuring strong flow diversion away from the aneurysm while keeping a low probability of occlusion of perforating vessels.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!