Background And Objective: Endoscopic cyclophotocoagulation is a laser modality used to treat refractory glaucoma through the ablation of ciliary epithelium on the ciliary processes, which results in reduced aqueous production. The purpose of this study was to investigate the effect of various viscoelastics and optimal probe positioning on endoscopic cyclophotocoagulation treatment.
Materials And Methods: Six different viscoelastic materials and a balanced salt solution, which was used as a control, were evaluated by placing 2 mL of each in a separate 2 mL glass tube. Laser power was measured at six distances to represent clinical placement of the laser probe in relation to the ciliary processes.
Results: The amount of energy measured with the probe at the bottom of the tube and 1 and 2 mm away was not statistically significantly different among the viscoelastic materials or the balanced salt solution. There was a gradual decrease in transmission of laser energy in all viscoelastics as the laser probe was moved farther away.
Conclusion: To preserve the intended laser energy setting, the optimal distance between the treated tissue and the laser probe is 2 mm.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3928/15428877-20080301-18 | DOI Listing |
Lasers Med Sci
November 2024
Shenzhen Eye Institute, Shenzhen Eye Hospital, Jinan University, Shenzhen, China.
Cureus
October 2024
Ophthalmology, Akasaka Shimazaki Eye Clinic, Tokyo, JPN.
Am J Ophthalmol Case Rep
December 2024
Casey Eye Institute, Oregon Health and Science University, Portland, OR, USA.
Purpose: The Hydrus microstent was approved by the FDA in August 2018 for use with cataract surgery to reduce IOP in patients with mild to moderate primary open angle glaucoma (POAG). Pivotal clinical trials demonstrated its overall safety and efficacy in lowering IOP. However, malpositioning of the implant can result in uveitis-glaucoma-hyphema (UGH) syndrome necessitating device explantation.
View Article and Find Full Text PDFAsia Pac J Ophthalmol (Phila)
June 2024
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Lab, Beijing 100730, China. Electronic address:
Purpose: To evaluate the long-term efficacy of endoscopic cyclophotocoagulation (ECP) via a pars plana approach in a large cohort of refractory glaucoma patients DESIGN: Single-center, retrospective, longitudinal, cohort study.
Methods: This study recruited patients who underwent ECP and consecutively visited and were followed up for at least 5 years at Beijing Tongren Eye Center, China from January 2013 to December 2017. All patients underwent a complete ophthalmic examination.
Indian J Ophthalmol
May 2024
Department of Ophthalmology, Faculty of Medicine Universitas Indonesia - Cipto Mangunkusumo Kirana Eye Hospital, Jakarta, Indonesia.
Age is an important risk factor for both glaucoma and cataract. As global life expectancy continues to rise, the prevalence of concomitant open-angle glaucoma (OAG) and cataracts is increasing. Currently, there is a lack of definitive consensus on the optimal management approach for such individuals.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!