Influence of CBCT exposure conditions on radiation dose.

Oral Surg Oral Med Oral Pathol Oral Radiol Endod

Craniofacial Imaging Center, Case Western Reserve University, School of Dental Medicine, Cleveland, OH 44106, USA.

Published: June 2008

Background: Cone-beam computed tomography (CBCT) has been changing the way dental practitioners use imaging. The radiation dose to the patient and how to effectively reduce the dose is still not completely clear to most users of this technology.

Objective: The objective of this study was to quantitate the change in radiation dose when using different CBCT settings.

Methods: A CBCT machine was modified to allow different setting combinations. The variables consisted of 4 different mA choices (2, 5, 10, and 15), 2 kVp choices (100 and 120), and 3 fields of view (6 inches, 9 inches, and 12 inches). A radiation phantom with 10 thermoluminescent dosimeters (TLD) was used to measure radiation dose. One specific setting (15 mA, 120 kVp, and 12-inch FOV) was scanned 3 times to determine consistency.

Results: The CBCT showed less than 5% variance in radiation dose values. An overall reduction in dose of about 0.62 times was achieved by reducing the kVp from 120 to 100. When reducing the field size the dose decreased 5% to 10%, while for organs that escaped the direct beam the reduction was far greater.

Conclusions: A reduction in radiation dose can be achieved by using the lowest exposure settings and narrow collimation.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.tripleo.2007.12.019DOI Listing

Publication Analysis

Top Keywords

radiation dose
24
dose
9
inches inches
8
radiation
7
influence cbct
4
cbct exposure
4
exposure conditions
4
conditions radiation
4
dose background
4
background cone-beam
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!