Characteristics of mechanical heart valve cavitation in a pneumatic ventricular assist device.

Artif Organs

Department of Artificial Organs, Research Institute, National Cardiovascular Center, Osaka, Japan.

Published: June 2008

In previous studies, we investigated the mechanism of mechanical heart valve (MHV) cavitation and cavitation intensity with a nonsynchronized experiment system. Our group is currently developing a pneumatic ventricular assist device (PVAD), and in this study we investigated MHV cavitation intensity in the PVAD using a synchronized analysis of the cavitation images and the acoustic signal of cavitation bubbles. A 23-mm Medtronic Hall valve with an opening angle of 70 degrees was mounted in the mitral position of the PVAD after removing the sewing ring. A function generator provided a square signal, which used the trigger signal of the electrocardiogram R wave (ECG-R) mode of the control-drive console for circulatory support. This square signal was delayed by a delay circuit and was then used as the trigger signal for a pressure sensor and a high-speed video camera. The data were stored using a digital oscilloscope at a 1-MHz sampling rate, and then the pressure signal was band-pass filtered between 35 and 200 kHz using a digital filter. The band-pass filtered root mean squared (RMS) pressure and cavitation cycle duration were used as an index of cavitation intensity. The cavitation bubbles were concentrated at the valve stop, and the cavitation cycle duration and RMS pressure increased as the heart rate and driving pressure increased. At the low valve-closing velocity, bubble cavitation was observed near the valve stop. However, at the fast valve-closing velocity, cloud cavitation was observed. A high-frequency signal wave was generated when the bubbles collapsed. The cavitation cycle duration and RMS pressure increased as the valve-closing velocity increased linearly.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1525-1594.2008.00564.xDOI Listing

Publication Analysis

Top Keywords

cavitation
13
cavitation intensity
12
rms pressure
12
cavitation cycle
12
cycle duration
12
pressure increased
12
valve-closing velocity
12
mechanical heart
8
heart valve
8
valve cavitation
8

Similar Publications

Focused Ultrasound and Microbubble-Mediated Delivery of CRISPR-Cas9 Ribonucleoprotein to Human Induced Pluripotent Stem Cells.

Mol Ther

January 2025

Department of Biology, Concordia University, 7141 Sherbrooke St. W H4B 1R6, Montreal, Canada; Department of Physics, Concordia University, 7141 Sherbrooke St. W H4B 1R6, Montreal, Canada. Electronic address:

CRISPR-Cas9 ribonucleoproteins (RNPs) have been heavily considered for gene therapy due to their high on-target efficiency, rapid activity and lack of insertional mutagenesis relative to other CRISPR-Cas9 delivery formats. Genetic diseases such as hypertrophic cardiomyopathy currently lack effective treatment strategies and are prime targets for CRISPR-Cas9 gene editing technology. However, current in-vivo delivery strategies for Cas9 pose risks of unwanted immunogenic responses.

View Article and Find Full Text PDF

The cavitation water jet cleaning and coating removal technique represents an innovative sustainable method for cleaning and removing coatings, with the nozzle serving as a crucial component of this technology. Developing an artificially submerged nozzle with a reliable structure and excellent cavitation performance is essential for enhancing cavitation water jets' cleaning and coating removal efficacy in an atmosphere environment (non-submerged state). This study is based on the shear flow cavitation mechanism of an angular nozzle, the resonance principle of an organ pipe, and the jet pump principle.

View Article and Find Full Text PDF

A 49-year old female, known to have had an asymptomatic pulmonary cavity since 2015, presented in 2023 with hemoptysis. Radiology showed a mass suspected to be an aspergilloma. Due to persistent hemoptysis, lobectomy was performed.

View Article and Find Full Text PDF

Pulsed Ultrasound-Mediated Drug Delivery Enhancement Through Human Sclera.

Transl Vis Sci Technol

January 2025

Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University, Shanghai, China.

Purpose: The purpose of this study was to characterize whether pulsed ultrasound (PUS) affects transscleral drug delivery.

Methods: Fluorescein sodium (NaF, 376 Da) and fluorescein isothiocyanate-conjugated dextran 40 (FD-40, 40 kDa) were used as model drugs. Human sclera grafts were placed in modified Franz diffusion cells and were treated by PUS (1 megahertz [MHz], 0.

View Article and Find Full Text PDF

Morphological Features Influence the Drug Loading and Delivery Efficacy of Photoactivatable Gold Nanocarriers for Antitumor Photo/Chemotherapy.

ACS Appl Mater Interfaces

January 2025

Institute of Optical Functional Materials for Biomedical Imaging, School of Chemistry and Pharmaceutical Engineering, Shandong First Medical University & Shandong Academy of Medical Science, Taian, Shandong 271016, PR China.

Photoactivatable gold nanocarriers are transforming antitumor therapies by leveraging their distinctive physicochemical properties, enabling targeted drug delivery and enhanced therapeutic efficacy in cancer treatment. This study systematically investigates how surface topography and morphology of gold nanocarriers influence drug loading capacity, light-to-heat conversion efficiency, and overall therapeutic performance in photo/chemotherapy. We synthesized four distinct morphologies of gold nanoparticles: porous gold nanocups (PAuNCs), porous gold nanospheres (PAuNSs), solid gold nanocups (SAuNCs), and solid gold nanospheres (SAuNSs).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!