The National Institute of Advanced Industrial Science and Technology (AIST) monopivot magnetic suspension centrifugal pump (MC105) was developed for open-heart surgery and several weeks of circulatory assist. The monopivot centrifugal pump has a closed impeller of 50 mm in diameter, supported by a single pivot bearing, and is driven through a magnetic coupling to widen the fluid gap. Design parameters such as pivot length and tongue radius were determined through flow visualization experiments, and the effectiveness was verified in preliminary animal experiments. The maximum overall pump efficiency reached 18%, and the normalized index of hemolysis tested with bovine blood was as low as 0.0013 g/100 L. Animal experiments with MC105 were conducted in sheep for 3, 15, 29, and 35 days in a configuration of left ventricle bypass. No thrombus was formed around the pivot bearing except when the pump speed was reduced by 20% of normal operational speed, which reduced the pump flow by 40% to avoid inlet suction. Subsequently, the antithrombogenic design was verified in animal experiments for 5 weeks at a minimum rotational speed of greater than 1500 rpm and a minimum pump flow greater than 1.0 L/min; no thrombus formation was observed under these conditions.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1525-1594.2008.00570.xDOI Listing

Publication Analysis

Top Keywords

centrifugal pump
12
animal experiments
12
monopivot magnetic
8
magnetic suspension
8
suspension centrifugal
8
circulatory assist
8
pivot bearing
8
speed reduced
8
pump flow
8
pump
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!