A polyacrylamide (PAAm)-modified monolithic silica capillary column of increased phase ratio, 200T-PAAm, for hydrophilic interaction liquid chromatography (HILIC) was prepared. The column showed high separation efficiency, with a theoretical plate height H = 7-20 microm at a linear velocity, u = 1-7 mm/s. From a kinetic plot analysis, it was expected that the monolithic column could provide three times faster separation than particle-packed HILIC columns under a pressure limit at 20 MPa. HILIC coupled with electrospray ionization (ESI)-mass spectrometry (HILIC-ESI-MS) using the 200T-PAAm column was employed for the analysis of underivatized carbohydrates to achieve fast and efficient separations of mixtures containing mono-, di-, and trisaccharides within 5 min. Under single MS full scan mode, 200 pg of oligosaccharides was detected by the system. The limit of detection (LOD) of the LC-ESI-MS/MS system was determined using selected reaction monitoring (SRM) to be as low as 3.2 ng/mL (attomol level) for nonreducing saccharides. The system was successfully applied to the detection of disaccharides in extracts of plant, such as corn, soybean, and Arabidopsis thaliana.