Objective: To compare monotonic biomechanical properties and fatigue life of a broad locking compression plate (LCP) fixation with a broad limited contact dynamic compression plate (LC-DCP) fixation to repair osteotomized equine third metacarpal (MC3) bones.

Study Design: In vitro biomechanical testing of paired cadaveric equine MC3 with a mid-diaphyseal osteotomy, stabilized by 1 of 2 methods for fracture fixation.

Animal Population: Cadaveric adult equine MC3 bones (n=12 pairs).

Methods: MC3 were divided into 3 groups (4 pairs each) for: (1) 4-point bending single cycle to failure testing; (2) 4-point bending cyclic fatigue testing; and (3) torsional single cycle to failure testing. The 8-hole, 4.5 mm LCP was applied to the dorsal surface of 1 randomly selected bone from each pair. One 8-hole, 4.5 mm LC-DCP) was applied dorsally to the contralateral bone from each pair. All plates and screws were applied using standard ASIF techniques. All MC3 bones had mid-diaphyseal osteotomies. Mean test variable values for each method were compared using a paired t-test within each group. Significance was set at P<.05.

Results: Mean yield load, yield bending moment, composite rigidity, failure load and failure bending moment, under 4-point bending, single cycle to failure, of the LCP fixation were significantly greater than those of the LC-DCP fixation. Mean cycles to failure for 4-point bending was significantly greater for the LCP fixation compared with LC-DCP fixation. Mean yield load, mean composite rigidity, and mean failure load under torsional testing, single cycle to failure was significantly greater for the broad LCP fixation compared with the LC-DCP fixation.

Conclusion: The 4.5 mm LCP was superior to the 4.5 mm LC-DCP in resisting the static overload forces (palmarodorsal 4-point bending and torsional) and in resisting cyclic fatigue under palmarodorsal 4-point bending.

Clinical Relevance: The results of this in vitro study may provide information to aid in the selection of a biological plate for the repair of equine long bone fractures.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1532-950X.2008.00378.xDOI Listing

Publication Analysis

Top Keywords

compression plate
16
vitro biomechanical
8
locking compression
8
plate fixation
8
dynamic compression
8
osteotomized equine
8
equine third
8
third metacarpal
8
equine mc3
8
mc3 bones
8

Similar Publications

The application of high-pressure grinding rolls (HPGR) for ore crushing is considered to be one of the effective ways to save energy and reduce emissions in the ore processing industry. The crushing effect is directly determined by the forces of ore material during roll crushing. However, the mechanical state of ore material in roll crushing and the effect of roll structure, process parameters, feed particle size, on the force during the crushing of ore material needs to be expanded.

View Article and Find Full Text PDF

Aims: The "2 to 10% strain rule" for fracture healing has been widely interpreted to mean that interfragmentary strain greater than 10% predisposes a fracture to nonunion. This interpretation focuses on the gap-closing strain (axial micromotion divided by gap size), ignoring the region around the gap where osteogenesis typically initiates. The aim of this study was to measure gap-closing and 3D interfragmentary strains in plated ovine osteotomies and associate local strain conditions with callus mineralization.

View Article and Find Full Text PDF

Functional Outcome of Subtrochanteric Femoral Fractures Fixation by Proximal Femoral Locking Compression Plate.

Mymensingh Med J

January 2025

Dr Md Sonaullah, Assistant Professor, Department of Orthopedics and Traumatology, Mymensingh Medical College (MMC), Mymensingh, Bangladesh; E-mail:

Subtrochanteric femoral fractures are one of the common fractures encountered in today's Orthopaedic practice. This area consists of mostly cortical bone with high stress generation thus heal slowly which leads implant failure. The inherent instability of this fracture and forces of the muscles with comminuted medial calcar is giving the fracture a tendency to varus collapse.

View Article and Find Full Text PDF

Prompt emergence from general anesthesia is crucial after neurosurgical procedures, such as craniotomies, to facilitate timely neurological evaluation for identification of intraoperative complications. Delayed emergence can be caused by residual anesthetics, metabolic imbalances, and intracranial pathology, for which an eye examination can provide early diagnostic clues. The sunset sign (or setting sun sign), characterized by a downward deviation of the eyes, can be an early indicator of raised intracranial pressure (ICP) or midbrain compression, as is commonly observed in states of hydrocephalus or periaqueductal or tectal plate dysfunction.

View Article and Find Full Text PDF

Background: The purpose of this study was to report the clinical and psychological outcomes of using a locking compression plate (LCP) as a sequential external fixator following the distraction phase in the treatment of tibial bone defects caused by fracture-related infection (FRI).

Methods: We retrospectively analyzed the clinical records and consecutive X-ray images of patients with tibial bone defects who were treated with an LCP as a sequential external fixator following the distraction phase, between June 2017 and December 2022. The ASAMI criteria were applied to assess the bone and functional outcomes, and postoperative complications were evaluated by using the Paley classification.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!