Aim: Feedback during CPR may facilitate quality in chest compressions, but has also been associated with caregiver complaints such as stiff wrists, the need for more force and increased fatigue. This concern about extra work is, when using the CPREzy with its own spring-loaded surface, particularly relevant in the face of an increased number of successive compressions. This manuscript evaluates the objective workloads for caregivers with and without the CPREzy.

Materials And Methods: An air pressure driven, piston device was used to generate controlled compressions in a manikin model. The pressure was applied for chest compressions with each of the following: the cylindrical end of the piston, a wooden block as dummy for the CPREzy, and the CPREzy itself. Three manikins with subjectively different spring compliances were selected for the tests. Series of 20 compressions were performed over a wide range of pressures.

Results: No additional force is required to achieve a given depth of compression with or without the CPREzy. However, some additional work is required, ranging from 21 to 26.5%. This work is caused by the longer compression distance associated with the need to compress two springs (e.g. the CPREzy and the chest wall) instead of one (e.g. the chest wall).

Conclusion: The subjective feeling of increased rescuer fatigue with the CPREzy can, at least in part, be attributed to the extra work required for compressing the spring of the CPREzy. Improved accuracy in chest compression depth is likely to be another, more significant, factor in rescuer fatigue.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.resuscitation.2008.01.024DOI Listing

Publication Analysis

Top Keywords

cprezy
8
chest compressions
8
extra work
8
work required
8
rescuer fatigue
8
work
5
chest
5
compressions
5
cprezy involve
4
involve work
4

Similar Publications

The impact of a CPRezy™ feedback device on the quality of chest compressions performed by nurses.

Am J Emerg Med

July 2018

Department of Emergency Medicine and Disaster, Medical University Bialystok, Bialystok, Poland; Department of Emergency Medicine, Medical University of Warsaw, Warsaw, Poland. Electronic address:

View Article and Find Full Text PDF

Background: Out-of-hospital cardiac arrest is a leading cause of mortality and serious neurological morbidity in Europe. We aim to investigate the effect of 3 cardiopulmonary resuscitation (CPR) feedback devices on effectiveness of chest compression during CPR.

Methods: This was prospective, randomized, crossover, controlled trial.

View Article and Find Full Text PDF

A randomised control trial of prompt and feedback devices and their impact on quality of chest compressions--a simulation study.

Resuscitation

April 2014

Academic Department of Critical Care, Anaesthesia and Pain, Heart of England NHS Foundation Trust, Birmingham B9 5SS, UK; University of Warwick, Warwick Medical School, CV4 7AL, UK. Electronic address:

Aim: This study aims to compare the effect of three CPR prompt and feedback devices on quality of chest compressions amongst healthcare providers.

Methods: A single blinded, randomised controlled trial compared a pressure sensor/metronome device (CPREzy), an accelerometer device (Phillips Q-CPR) and simple metronome on the quality of chest compressions on a manikin by trained rescuers. The primary outcome was compression depth.

View Article and Find Full Text PDF

Aim: Feedback during CPR may facilitate quality in chest compressions, but has also been associated with caregiver complaints such as stiff wrists, the need for more force and increased fatigue. This concern about extra work is, when using the CPREzy with its own spring-loaded surface, particularly relevant in the face of an increased number of successive compressions. This manuscript evaluates the objective workloads for caregivers with and without the CPREzy.

View Article and Find Full Text PDF

Monitoring the quality of cardiopulmonary resuscitation.

Curr Opin Crit Care

June 2007

Intensive Care Unit, Royal Melbourne Hospital, Parkville, Victoria, Australia.

Purpose Of Review: Numerous recent reports have described limitations in the quality of cardiopulmonary resuscitation. Thus, there has been increasing interest in the techniques available to monitor quality. This review focuses on the major publications since the review published by the International Liaison Committee on Resuscitation in 2005.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!