Evaluation of a novel ventricular support device with defibrillation capabilities in canine and porcine animal models.

J Cardiovasc Electrophysiol

Cardiac Rhythm Management Laboratory, Division of Cardiovascular Diseases, Department of Medicine, University of Alabama at Birmingham, AL 35294-0019, USA.

Published: August 2008

Introduction: Sudden death is prevalent in heart failure patients. We tested an implantable ventricular support device consisting of a wireform harness with one or two pairs of integrated defibrillation electrode coils.

Methods And Results: The device was implanted into six pigs (36-44 kg) through a subxiphoid incision. Peak voltage (V) defibrillation thresholds (DFT) were determined for five test configurations compared with a control transvenous lead (RV to CanPect). Defibrillator can location (abdominal or pectoral) and common coil separation on the implant (0 degrees or 60 degrees ) were studied.(.) The DFT for RV60 to LV60 + CanPect was significantly less than control (348 +/- 57 vs 473 +/- 27 V, P < 0.05). The DFTs for other vectors were similar to control except for RV0 to LV0 + CanAbd (608 +/- 159 V). The device was implanted into 12 adult dogs for 42, 90, or 180 days with DFT and pathological examination performed at the terminal study. Cardiac pressures were determined at baseline, after implantation, and at the terminal study. The DFT was also determined in a separate group of four dogs at 42 days following implantation of the support device with one pair of defibrillation electrodes. The DFTs at implant and explant in dogs with one pair (8 +/- 1.5 Joules [J] and 6 +/- 1.9 J) or two pairs (8 +/- 3.4 J and 7 +/- 1.9 J) of defibrillation electrodes were not significantly different from each other but significantly less than control measured at the terminal study (18 +/- 3.4 J). Left-sided pressures were significantly decreased at explant but within expected normal ranges. Right-sided pressures were not different except for RV systolic. Histopathology indicated mild to moderate epicardial inflammation and fibrosis, consistent with a foreign body healing response.

Conclusions: This defibrillation-enabled ventricular support system maintained mechanical functionality for up to 6 months while inducing typical chronic healing responses. The DFT was equal to or lower than a standard transvenous vector.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8167.2008.01132.xDOI Listing

Publication Analysis

Top Keywords

ventricular support
12
support device
12
terminal study
12
device implanted
8
dft determined
8
+/-
8
defibrillation electrodes
8
device
5
defibrillation
5
dft
5

Similar Publications

ECLS-SHOCK and DanGer Shock: Implications for optimal temporary mechanical circulatory support use for cardiogenic shock due to acute myocardial infarction.

Can J Cardiol

January 2025

University of Montreal Hospital Center (CHUM) Cardiovascular Center & Research Center (CRCHUM), University of Montreal, Montreal, Quebec, Canada. Electronic address:

Despite concerted efforts to rapidly identify patients with cardiogenic shock complicating acute myocardial infarction (AMI-CS) and provide timely revascularization, early mortality remains stubbornly high. While artificially augmenting systemic flow through the use of temporary mechanical circulatory support (tMCS) devices would be expected to reduce the rate of progression to multi-organ dysfunction and thereby enhance survival, reliable evidence for benefit has remained elusive with lingering questions regarding the appropriate selection of both patients and devices, as well as the timing of device implantation relative to other critical interventions. Further complicating matters are the resource-intensive multidisciplinary systems of care that must be brought to bear in this complex patient population.

View Article and Find Full Text PDF

Outcomes after medical treatment for primary aldosteronism: an international consensus and analysis of treatment response in an international cohort.

Lancet Diabetes Endocrinol

January 2025

Centre for Endocrinology and Metabolism, Hudson Institute of Medical Research, Clayton, VIC, Australia; Department of Endocrinology, Monash Health, Clayton, VIC, Australia.

Background: Primary aldosteronism can be treated medically but there is no standardised method to evaluate treatment outcomes. We aimed to develop criteria for assessing the outcomes of targeted medical treatment of primary aldosteronism, analyse outcomes across an international cohort, and identify factors associated with a complete treatment response.

Methods: An international panel of 31 primary aldosteronism experts used the Delphi method to reach consensus on the definition of complete, partial, or absent biochemical and clinical outcomes of medical treatment of primary aldosteronism.

View Article and Find Full Text PDF

Single cell approaches define neural stem cell niches and identify microglial ligands that can enhance precursor-mediated oligodendrogenesis.

Cell Rep

January 2025

Michael Smith Laboratories, University of British Columbia, Vancouver, BC V6T 1Z4, Canada; Department of Medical Genetics, University of British Columbia, Vancouver, BC V6T 1Z3, Canada; Program in Neurosciences and Mental Health, Hospital for Sick Children, Toronto, ON M5G 0A4, Canada; Institute of Medical Science, University of Toronto, Toronto, ON M5S 1A8, Canada; Department of Molecular Genetics, University of Toronto, Toronto, ON M5S 1A8, Canada. Electronic address:

Here, we used single cell RNA sequencing and single cell spatial transcriptomics to characterize the forebrain neural stem cell (NSC) niche under homeostatic and injury conditions. We defined the dorsal and lateral ventricular-subventricular zones (V-SVZs) as two distinct neighborhoods and showed that, after white matter injury, NSCs are activated to make oligodendrocytes dorsally for remyelination. This activation is coincident with an increase in transcriptionally distinct microglia in the dorsal V-SVZ niche.

View Article and Find Full Text PDF

Background: Fulminant myocarditis (FM) is a potentially lethal disease with a wide spectrum of clinical presentation, thus making the diagnosis hard to depict. In cases where acute circulatory failure occurs venoarterial (VA) extracorporeal membrane oxygenation (ECMO) support is a valid management strategy, especially in the pediatric and adult patients. This study aims to report the results of VA ECMO for FM in our Institution.

View Article and Find Full Text PDF

TEER for SAM of the Mitral Valve and Flail Posterior Mitral Leaflet: 1-Clip Solution.

JACC Case Rep

January 2025

Department of Cardiovascular Medicine, Richmond Heart & Vascular Associates, Richmond, Virginia, USA.

Transcatheter edge-to-edge repair (TEER) is approved for patients with symptomatic severe mitral regurgitation (MR) who are deemed inoperable or at high surgical risk with life expectancy of more than 1 year, but has also been used off-label in patients with hypertrophic obstructive cardiomyopathy (HOCM) for symptomatic relief who are not candidates for septal reduction therapy. An 83-year-old woman with decompensated heart failure was found to have HOCM with systolic anterior motion of the mitral valve and a large P2 flail segment with ruptured cords. TEER was performed resulting in mild MR and resolution of the prior left ventricular outflow tract gradient.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!