A new instrument to measure the in situ bidirectional reflectance distribution function (BRDF) of surfaces is described. This instrument measures the BRDF for eight illumination angles from 0 to 65 deg, three colors (475, 570, and 658 nm), and at over 100 selected viewing angles. The viewing zenith angles range from 5 to 65 deg, and the azimuth angles, relative to the illumination direction, range from 0 to ?180 deg. Many tests of the system have been run and show that for flat surfaces the BRDF of a sample surface can be measured with a precision of 1-5% and an accuracy of 10% of the measured reflectance. The BRDF for a dry and wet sand sample is presented as a demonstration of the instrument.

Download full-text PDF

Source
http://dx.doi.org/10.1364/ao.39.006197DOI Listing