Calibration of a 300-mm-aperture phase-shifting Fizeau interferometer.

Appl Opt

Telecommunications and Industrial Physics, Commonwealth Scientific and Industrial Research Organisation, P.O. Box 218, Lindfield 2070, New South Wales, Australia.

Published: October 2000

A 300-mm-aperture digital phase-shifting Fizeau interferometer has been developed in-house for precision metrology of optical components fabricated by the optical workshop at Telecommunications and Industrial Physics, Commonwealth Scientific and Industrial Research Organization. We describe the procedures used in the calibration of the instrument. A reference data file representing the deviations from flatness of the reference surface is generated, measurement uncertainty estimated, and aberrations in the instrument assessed. Measurements on 250-mm-diameter uncoated optical surfaces have consistently shown short-term repeatability of 0.3-nm rms from measurement to measurement and allowed for absolute characterization of these surfaces to within a few nanometers.

Download full-text PDF

Source
http://dx.doi.org/10.1364/ao.39.005161DOI Listing