Human subject testing of leakage in a loose-fitting PAPR.

J Occup Environ Hyg

Fischell Department of Bioengineering, University of Maryland, College Park, Maryland 20742, USA.

Published: May 2008

Leakage from loose-fitting PAPRs (powered air-purifying respirators) can compromise the safety of wearers. The Martindale Centurion MAX multifunction PAPR is a loose-fitting PAPR that also incorporates head, eye, and ear protection. This respirator is used in mines where coal dust usually is controlled by ventilation systems. Should the respirator be depended on for significant respiratory protection? Ten human volunteers were asked to wear the Centurion MAX inside a fog-filled chamber. Their inhalation flow rates were measured with small pitot-tube flowmeters held inside their mouths. They were video imaged while they breathed deeply, and the points at which the fog reached their mouths were determined. Results showed that an average of 1.1 L could be inhaled before contaminated air reached the mouth. As long as the blower purges contamination from inside the face piece during exhalation, the 1.1 L acts as a buffer against contaminants leaked due to overbreathing of blower flow rate.

Download full-text PDF

Source
http://dx.doi.org/10.1080/15459620801996819DOI Listing

Publication Analysis

Top Keywords

leakage loose-fitting
8
loose-fitting papr
8
centurion max
8
human subject
4
subject testing
4
testing leakage
4
papr leakage
4
loose-fitting paprs
4
paprs powered
4
powered air-purifying
4

Similar Publications

8-hour performance of loose-fitting powered air-purifying respirators in simulated hospital and coal mine environments.

J Occup Environ Hyg

December 2023

Key Laboratory of Coal Methane and Fire Control, Ministry of Education, China University of Mining and Technology, Xuzhou, Jiangsu, China.

Loose-fitting powered air-purifying respirators (LF-PAPRs) are increasingly used in hospitals and coal mines because of their high comfort and protection level, but the utilization faces the challenges of 8-hr continuous high protection requirements in the hospital environment and the coupling effects of high temperature, high humidity, high dust concentration in coal mines. Based on the self-developed powered air-purifying respirator simulation test system, this study explores the 8-hr changes of supplied airflow, the relative air pressure inside the inlet covering (ΔP), and total inward leakage (TIL) of four models of LF-PAPRs in simulated hospital and coal mine environments. Results show that: (1) In a simulated hospital environment, all four LF-PAPRs showed filter cartridge blockage within 5 ∼ 6 hr of continuous operation; while in the simulated coal mine, three models of LF-PAPRs showed filter cartridge blockage within 3 hr.

View Article and Find Full Text PDF

Loose-fitting powered air-purifying respirators (PAPRs) are used in healthcare settings, although barriers to routine, everyday usage remain, including usability concerns and potential interference with work activities. Loose-fitting PAPRs are approved by the National Institute for Occupational Safety and Health (NIOSH) and must meet minimum performance requirements, including a minimum airflow requirement of 170 L/min. One course of action to address usability concerns is to allow for the use of PAPRs designed with reduced airflow rates.

View Article and Find Full Text PDF

Recently, total inward leakage (TIL) for filtering facepiece and elastomeric half-mask respirators (EHRs) was measured according to the International Organization for Standardization (ISO) test method standard 16900-1:2014 that showed larger TIL for corn oil aerosol than for NaCl aerosol. Comparison of TIL measured for different aerosols for higher protection level respirators is lacking. The objective of this study was to determine TIL for EHRs, full-facepiece respirators, and loose-fitting and tight-fitting powered air-purifying respirators (PAPRs) using NaCl and corn oil aerosols to compare.

View Article and Find Full Text PDF

Objectives: N95 filtering facepiece respirators (N95 FFRs) and surgical masks are comprised of multiple layers of nonwoven polypropylene. Tight-fitting N95 FFRs are respiratory protective devices (RPDs) designed to efficiently filter aerosols. During the COVID-19 pandemic, health care workers (HCWs) throughout the world continue to face shortages of disposable N95 FFRs.

View Article and Find Full Text PDF

In surgical settings, infectious particulate wound contamination is a recognized cause of post-operative infections. Powered air purifying respirators (PAPRs) are worn by healthcare workers for personal protection against contaminated aerosols. Healthcare infection preventionists have expressed concern about the possibility that infectious particles expelled from PAPR exhalation channels could lead to healthcare-associated disease, especially in operative settings where sterile procedural technique is essential.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!