This paper presents the first microneedle-based transdermal patch with integrated active dispensing functionality. The electrically controlled system consists of a low-cost dosing and actuation unit capable of controlled release of liquid in the microliter range at low flow-rates and minimally invasive, side-opened, microneedles. The system was successfully tested in vivo by insulin administration to diabetic rats. Active infusion of insulin at 2 mul/h was compared to passive, diffusion-driven, delivery. Continuous active infusion caused significantly higher insulin concentrations in blood plasma. After a 3-h delivery period, the insulin concentration was five times larger compared to passive delivery. Consistent with insulin concentrations, actively administered insulin resulted in a significant decrease of blood glucose levels. Additionally, insertion and liquid injection was verified on human skin. This study shows the feasibility of a patch-like system with on-board liquid storage and dispensing capability. The proposed device represents a first step towards painless and convenient administration of macromolecular drugs such as insulin or vaccines.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1109/TBME.2007.906492 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!