Unlabelled: In this study, we evaluated clinical images to determine appropriate settings for slice thickness in screening examinations for hepatocellular carcinoma and metastatic liver cancer, which are frequently performed in abdominal CT. The clinical images of 15 cases screened for hepatocellular carcinoma and 15 cases screened for metastatic liver cancer were evaluated. The evaluation was visually performed by three abdominal radiologists, and the sensitivity and specificity of each slice thickness were calculated.

Results: differences in sensitivity and specificity were not found between slice thicknesses of 2.5 mm and 5.0 mm; however, sensitivity and specificity were low, and confidence was also low at a 10.0 mm slice thickness. Furthermore, when a 5.0 mm slice thickness was adopted, it was shown that radiation dose in limited parts could be reduced greatly to a noise level that compared equally with a study in which a 2.5 mm slice thickness was adopted in another evaluation. Therefore, the objective could be achieved by using a slice thickness of about 5.0 mm in multidetector-row CT examinations for the screening of liver cancer, while controlling patient dose to a minimum.

Download full-text PDF

Source
http://dx.doi.org/10.6009/jjrt.64.65DOI Listing

Publication Analysis

Top Keywords

slice thickness
28
liver cancer
12
sensitivity specificity
12
slice
8
multidetector-row examinations
8
examinations screening
8
screening liver
8
clinical images
8
hepatocellular carcinoma
8
metastatic liver
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!