Influence on fluid dynamics of coronary artery outlet angle variation in artificial aortic root prosthesis.

Biomed Eng Online

MVIP ImagingProducts GmbH, Nörten-Hardenberg, Germany.

Published: February 2008

AI Article Synopsis

Article Abstract

Background: Because of higher life expectancy, the number of elderly patients today with degenerative aortic diseases is on the increase. Often artificial aortic roots are needed to replace the native tissue. This surgical procedure requires re-implantation of the previous separated coronary arteries into the wall of the prosthesis. Regardless of the prosthesis type, changes in the reinsertion technique, e.g., the variation of the outlet angle of the coronary arteries, could influence the coronary blood flow. Whether the prosthesis type or the outlet angle variation significantly improves the blood circulation and lowers the risk of coronary insufficiency is still an open question. The numerical calculations presented can help to clear up these disputable questions.

Methods: Two simplified base geometries are used for simulating the blood flow in order to determine velocity and pressure distributions. One model uses a straight cylindrical tube to approximate the aortic root geometry; the other uses a sinus design with pseudosinuses of Valsalva. The coronary outlet angle of the right coronary artery was discretely modified in both models in the range from 60 degrees to 120 degrees . The pressure and velocity distributions of both models are compared in the ascending aorta as well as in the right and the left coronary artery.

Results: The potentially allowed and anatomic limited variation of the outlet angle influences the pressure only a little bit and shows a very slight relative maximum between 70 degrees and 90 degrees . The sinus design and variations of the outlet angle of the coronary arteries were able to minimally optimize the perfusion pressure and the velocities in the coronary circulation, although the degree of such changes is rather low and would probably not achieve any clinical influence.

Conclusion: Our results show that surgeons should feel relatively free to vary the outlet angle within the anatomic structural conditions when employing the technique of coronary reinsertion.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2275278PMC
http://dx.doi.org/10.1186/1475-925X-7-9DOI Listing

Publication Analysis

Top Keywords

outlet angle
28
coronary arteries
12
angle coronary
12
coronary
11
coronary artery
8
angle variation
8
artificial aortic
8
aortic root
8
prosthesis type
8
variation outlet
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!