Aims: TTK Chitra heart valve prosthesis (CHVP), a tilting disc mechanical heart valve of low cost and proven efficacy, has been in use for the last 15 years. Although various studies substantiating its long-term safety and efficacy are available, no study had assessed its echocardiographic characteristics. The purpose of this study was to determine the normal Doppler parameters of CHVP in the mitral position and to assess whether derivation of mitral valve area (MVA) using the continuity equation (CE) and more commonly used pressure half-time (PHT) method is comparable in the functional assessment of this tilting disc mitral prosthesis.

Methods And Results: Doppler echocardiography was performed in 40 consecutive patients with CHVP in the mitral position. All patients were clinically stable, without evidence of prosthetic valve dysfunction such as significant obstruction or regurgitation, endocarditis, left ventricular dysfunction (ejection fraction <40%), or significant aortic regurgitation. Valve sizes studied included 25, 27, and 29 mm. Mitral valve area was derived both by the PHT method and by the CE, using the stroke volume measured in the ventricular outflow tract divided by the time-velocity integral of CHVP jet. The peak Doppler gradient ranged from 5 to 21 (mean 11.0) mmHg, and the mean gradient ranged from 1.7 to 9.2 (mean 4.1) mmHg. Mean gradient negatively correlated with an increase in the actual orifice area (AOA) derived from the valve orifice diameter given by the manufacturer (r = -0.45, P = 0.004). Mitral valve area calculated by both PHT and CE increased significantly with an increase in the AOA (r = 0.42, P = 0.007 and r = 0.32, P = 0.046, respectively). Mitral valve area by the CE averaged 1.55 +/- 0.36 cm(2) (range 0.85 cm(2) for a 25 mm valve to 2.41 cm(2) for a 29 mm valve) and was smaller than by PHT (mean 2.04 +/- 0.41 cm(2), range 1.40-3.14 cm(2); P = 0.0001; t-test), irrespective of whether PHT is less than or >110 ms.

Conclusion: The Doppler parameters obtained with CHVP in the mitral position are comparable with those obtained with the different prosthetic valves in common use. In the selected group of patients with CHVP, assessment of MVA by the PHT method is comparable with that by the CE. Areas by both methods were smaller than the AOA provided by the manufacturer, as seen in other similar design valves.

Download full-text PDF

Source
http://dx.doi.org/10.1093/ejechocard/jen001DOI Listing

Publication Analysis

Top Keywords

mitral position
16
heart valve
12
chvp mitral
12
ttk chitra
8
tilting disc
8
doppler parameters
8
parameters chvp
8
pht method
8
method comparable
8
patients chvp
8

Similar Publications

We report a 75-year-old female with a history of two heart operations: aortic valve replacement (St. Jude Medical 21 mm) at the age of 44 years for severe rheumatic aortic stenosis and mitral valve replacement (Carbomedics 29 mm) at the age of 51 years for rheumatic mitral regurgitation. Decades later, she presented with exertional dyspnea.

View Article and Find Full Text PDF

The need for a permanent pacemaker (PPM) implantation after surgical aortic valve implantation (SAVR) is a recognized postoperative complication, with potentially long-term reduced survival. From 1987 to 2017, 2500 consecutive patients underwent SAVR with a biological valve with or without concomitant procedures such as CABG or mitral valve repair. Mechanical valves or valves in another position were excluded.

View Article and Find Full Text PDF

Background: The Edwards Sapien percutaneous valve (Edwards Lifesciences, Irvine, CA, USA) is a promising therapeutic option for congenital atrioventricular disease mostly because of the possibility to accommodate somatic growth with balloon dilatation.

Case Summary: This article reports the performance of the Edwards Sapien 3 valve in atrioventricular valve position in four paediatric patients.

Discussion: Despite aggressive antiplatelet and anticoagulation strategies, most patients showed early bioprosthesis dysfunction, with increasing gradient not related with somatic growth.

View Article and Find Full Text PDF

In recent years, the use of transcatheter valve-in-valve implantation in the mitral position (TMVI) for the treatment of mitral valve pathology following ring or bioprosthetic implantation has emerged as a less invasive option in comparison to repeated mitral valve surgery (RMVS). We aimed to compare the early and mid-term results of these two strategies. We retrospectively analyzed all patients who underwent a mitral intervention in our institution between 2005 and 2022.

View Article and Find Full Text PDF

Merging mixed reality and computational modeling for enhanced visualization of cardiac biomechanics.

Med Eng Phys

December 2024

Department of Engineering, Università degli Studi di Palermo, Viale delle Scienze Ed.8, Palermo, Italy; Department of Research, IRCCS ISMETT via Tricomi, 5, Palermo, Italy. Electronic address:

Mixed reality (MR) has the potential to complement numerical simulations for enhanced post-processing and integrate digital models into the daily clinical practice of healthcare professionals. In complex cardiac anatomies, the decision-making process for bioprosthesis implantation involves the challenging analysis of heart valve distribution, positioning, and sealing. This study proposes a framework to visualize computational modeling results in an immersive environment for comprehensive analysis of the geometric implications of implanted devices on human heart function.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!