Background: We here describe the pharmacological characteristic, in vivo efficacy, and in vitro mechanisms of a polymer-free leflunomide eluting stent in comparison to its rapamycin-coated equivalent.

Methods: Stents were coated with 40 mM solutions of leflunomide (L) or rapamycin (R) or were left uncoated (BM). Neointima formation was assessed 6 weeks after implantation into Sprague Dawley rats by optical coherence tomographies (OCT) and histopathology. In vitro proliferation assays were performed using isolated endothelial and smooth-muscle-cells from Sprague Dawley rats to investigate the cell-specific pharmacokinetic effect of leflunomide and rapamycin.

Results: HPLC-based drug release kinetics revealed a similar profile with 90% of the drug being released after 12.1+/-0.2 (L) and 13.0+/-0.2 days (R). After 6 weeks, OCTs showed that in-stent luminal obliteration was less for the coated stents (L:12.0+/-9.4%, R:13.3+/-13.1%) when compared to identical bare metal stents (BM:26.4+/-4.7%; p
Conclusions: The diminished inhibition of EC proliferation may improve arterial healing and contribute to the safety profile of the leflunomide stent.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.atherosclerosis.2007.12.055DOI Listing

Publication Analysis

Top Keywords

polymer-free leflunomide
8
leflunomide eluting
8
eluting stent
8
sprague dawley
8
dawley rats
8
introducing polymer-free
4
leflunomide
4
stent background
4
background describe
4
describe pharmacological
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!