The present study aimed to compare two commercially available actigraphs, with a concurrent polysomnographic (PSG) recording. Twelve healthy volunteers (six women; age range 19-28 yrs) simultaneously wore the Basic Mini-Motionlogger and Actiwatch for seven overnight polysomnographic recordings. Comparisons of the following sleep measures were focused on: sleep onset latency (SOL), total sleep time, wake after sleep onset, and sleep efficiency. Both devices underestimated SOL in comparison to PSG, but they had similar performance compared to PSG for the other sleep measures. A limit of the study is that the results can be only generalized to healthy young subjects.

Download full-text PDF

Source
http://dx.doi.org/10.1080/07420520801897228DOI Listing

Publication Analysis

Top Keywords

healthy young
8
young subjects
8
sleep measures
8
sleep onset
8
sleep
6
comparison actigraphs
4
actigraphs polysomnography
4
polysomnography healthy
4
subjects study
4
study aimed
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?