We describe the application of a Bayesian variable-number sample-path (VNSP) optimization algorithm to yield a robust design for a floating sleeve antenna for hepatic microwave ablation. Finite element models are used to generate the electromagnetic (EM) field and thermal distribution in liver given a particular design. Dielectric properties of the tissue are assumed to vary within +/- 10% of average properties to simulate the variation among individuals. The Bayesian VNSP algorithm yields an optimal design that is a 14.3% improvement over the original design and is more robust in terms of lesion size, shape and efficiency. Moreover, the Bayesian VNSP algorithm finds an optimal solution saving 68.2% simulation of the evaluations compared to the standard sample-path optimization method.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0031-9155/53/4/016DOI Listing

Publication Analysis

Top Keywords

sleeve antenna
8
antenna hepatic
8
hepatic microwave
8
microwave ablation
8
bayesian vnsp
8
vnsp algorithm
8
design
5
design optimization
4
optimization robust
4
robust sleeve
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!