Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: The combined measurement of diffusing capacity of the lung for nitric oxide (Dlno) and diffusing capacity of the lung for carbon monoxide (Dlco) is a simple, noninvasive tool, but methodologic factors might influence results and reproducibility. We thus quantified the influence of breath-hold time on Dlco and Dlno in subjects with or without airway disease.
Methods: Simultaneous single-breath measurements of Dlco and Dlno were performed in 10 patients with cystic fibrosis (CF) [mean +/- SD age, 33 +/- 9 years; FEV(1), 69 +/- 28% of predicted] and 10 healthy subjects (age, 31 +/- 9 years; FEV(1), 108 +/- 8% of predicted), using the Masterscreen PFT (Viasys/Jaeger; Höchberg, Germany), with 45 ppm of inspired nitric oxide (NO), and breath-hold times of 4 s, 6 s, 8 s, and 10 s. The last two of three consecutive measurements were used for analysis.
Results: In healthy subjects but not patients with CF, Dlno, and Dlco differed significantly (p < 0.05 each) between breath-hold times. Differences primarily occurred at 4 s and 10 s, while at 6 s and 8 s alveolar volume (VA), Dlno, Dlco, and Dlno/Dlco were similar. Variability of consecutive measurements (either three or the last two measurements) did not depend on breath-hold time. At 8 s, mean variabilities of Dlno and Dlco in healthy subjects were 4.9% and 2.5%, respectively, and 4.2% and 3.2% at 6 s. At 8 s, mean variabilities of Dlno and Dlco in CF patients were 4.4% and 1.9%, and 7.4% and 3.3% at 6 s.
Conclusions: Single-breath determinations of dlno and dlco showed no difference between breath-hold times of 6 s and 8 s in subjects with or without airway obstruction, and reproducibility was acceptable. Standardization of breath-hold time for Dlno measurements seems important for clinical and research comparisons.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1378/chest.07-2388 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!