Percutaneous treatment for mitral regurgitation: the QuantumCor system.

J Interv Cardiol

St. Luke's Medical Center, University of Arizona, College of Medicine, Phoenix, Arizona 85006, USA.

Published: April 2008

Aims: Percutaneous edge-to-edge techniques and annuloplasty have been used to treat mitral regurgitation (MR). However, neither intervention can be performed reliably a second time and, with annuloplasty, a foreign body is left behind. The mitral and tricuspid annuli are areas of dense collagen (Fig. 1); treatment with radiofrequency (RF) energy in sheep reduces their size, and can be repeated without affecting the coronary sinus. RF energy may also be used in leaflet procedures. Our aim was to improve mitral valve competence using techniques that can be incorporated into a minimally invasive approach.

Methods: In open-heart procedures in 16 healthy sheep (6 with naturally occurring MR), we used a malleable probe (QuantumCor, Inc., Lake Forest, CA) that conforms to the annular shape to deliver RF energy via a standard generator to replicate a surgical mitral annular ring. Four segments of the posterior mitral valve annulus were treated while on cardiopulmonary support via a left thoracotomy with access via the atrial appendage. Seven sheep were followed chronically.

Results: All sheep underwent intracardiac echocardiography (ICE) or direct circumferential measurement of the mitral annulus before and after RF therapy. RF therapy was administered in less than 4 minutes in each case, and the mean anteroposterior (AP) annular distance was reduced by a mean of 5.75 +/- 0.86 mm (23.8% reduction, P< 0.001). In the 6 sheep with nonischemic MR, regurgitation was eliminated. Acute histopathology (HP) demonstrated no damage to the leaflets, coronary sinuses, or coronary arteries. At 30 days, the AP distance continued to be reduced in the 7 surviving sheep (mean 5.0 +/- .6 mm, 21.4% reduction, P< 0.001).

Conclusions: In a sheep model, RF energy applied for less than 4 minutes per case at subablative temperatures in four quadrants of the posterior mitral valve annulus reduced the AP and circumferential annular distances significantly, and eliminated nonischemic MR. Results will need to be confirmed in follow-up studies to determine safety and efficacy. RF energy administered as a novel, percutaneous method of mitral valve annuloplasty may have the potential to reduce morbidity and mortality associated with current surgical techniques.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8183.2007.00333.xDOI Listing

Publication Analysis

Top Keywords

mitral valve
16
mitral
9
mitral regurgitation
8
posterior mitral
8
valve annulus
8
minutes case
8
sheep
7
energy
5
percutaneous treatment
4
treatment mitral
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!