A 3.7 MHz phased array probe using 0.91Pb(Zn(1/3)Nb(2/3))O(3)-0.09PbTiO(3 ).

IEEE Trans Ultrason Ferroelectr Freq Control

Research Laboratories, Research and Development Center, Toshiba Corporation, Kawasaki 210-8582, Japan.

Published: December 2009

A novel 128-channel phased array probe for echocardiography with a center frequency of 3.7 MHz using 0.91Pb(Zn(1/3)Nb(2/3))O(3)-0.09PbTiO(3 ) (PZN-9%PT) single crystal has been fabricated to realize greater sensitivity and broader bandwidth properties. The echo amplitude of the PZN-9%PT single-crystal probe is about 5 dB higher than that of the conventional lead airconate titanate (PZT) ceramic probe, and the fractional bandwidth is about 25 percentage points broader. The quality of B mode images obtained by the PZN-9%PT probe satisfies the performance of the two types of conventional PZT ceramic probes that have center frequencies of 2.5 and 3.75 MHz. At the reference frequency of 3 MHz, the Doppler sensitivity of the PZN-9%PT probe is about 5 dB higher than that of the 3.75 MHz PZT probe; the blood flow of a pulmonary vein in a hard-to-image patient is much more clearly imaged than in the case of using the PZT probe. These superior images are attributable to the use of sufficiently large PZN-9%PT single crystals obtained by the self-flux method.

Download full-text PDF

Source
http://dx.doi.org/10.1109/58.753031DOI Listing

Publication Analysis

Top Keywords

phased array
8
probe
8
array probe
8
frequency mhz
8
pzn-9%pt single
8
probe higher
8
pzt ceramic
8
pzn-9%pt probe
8
375 mhz
8
pzt probe
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!