Background: The IFCC Reference Measurement System for hemoglobin (Hb)A(1c) (IFCC-RM) has been developed within the framework of metrologic traceability and is embedded in a network of 14 reference laboratories. This paper describes the outcome of 12 intercomparison studies (periodic evaluations to control essential elements of the IFCC-RM).
Methods: Each study included: unknown samples (to test individual network laboratories); known samples (controls); recently manufactured calibrators (to check calculated assigned value); stored calibrators (to test stability) and a calibration-set (to calibrate the IFCC-RM). The unknown samples are measured by use of the IFCC-RM and the designated comparison methods [DCMs; the National Glycohemoglobin Standardization Program (NGSP) in the US, Japanese Diabetes Society/Japanese Society for Clinical Chemistry (JDS/JSCC) in Japan, and Mono-S in Sweden] are used to investigate the stability of the Master Equation (ME), the relationship between IFCC-RM and DCMs.
Results: A total of 105 IFCC-RM data sets were evaluated: 95 were approved, 5 were not, and for 5 no data were submitted. Trend analysis of the MEs, expressed as change in percentage HbA(1c) per year, revealed 0.000% (NGSP, not significant), -0.030%, (JDS/JSCC; significant) and -0.016% (Mono-S; not significant). Evaluation of long-term performance revealed no systematic change over time; 2 laboratories showed significant bias, 1 poor reproducibility. The mean HbA(1c) determined by laboratories performing mass spectrometry (MS) was the same as the mean determined by laboratories using capillary electrophoresis (CE), but the reproducibility at laboratories using CE was better. One batch of new calibrators was not approved. All stored calibrators were stable.
Conclusion: A sound reference system is in place to ensure continuity and stability of the analytical anchor for HbA(1c).
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1373/clinchem.2007.097402 | DOI Listing |
EJIFCC
December 2024
National Reference Laboratory, Abu Dhabi, UAE.
Background: An increasing number of wearable medical devices are being used for personal monitoring and professional health care purposes. These mobile health devices collect a variety of biometric and health data but do not routinely connect to a patient's electronic health record (EHR) or electronic medical record (EMR) for access by a patient's health care team.
Methods: The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Committee on Mobile Health and Bioengineering in Laboratory Medicine (C-MHBLM) developed consensus recommendations for consideration when interfacing mobile health devices to an EHR/EMR.
Anal Bioanal Chem
January 2025
Chemical Sciences Division, National Institute of Standards and Technology (NIST), Gaithersburg, MD, 20899, USA.
Commutability is where the measurement response for a reference material (RM) is the same as for an individual patient sample with the same concentration of analyte measured using two or more measurement systems. Assessment of commutability is essential when the RM is used in a calibration hierarchy or to ensure that clinical measurements are comparable across different measurement procedures and at different times. The commutability of three new Standard Reference Materials (SRMs) for determining serum total 25-hydroxyvitamin D [25(OH)D], defined as the sum of 25-hydroxyvitamin D [25(OH)D] and 25-hydroxyvitamin D [25(OH)D], was assessed through an interlaboratory study.
View Article and Find Full Text PDFClin Chem Lab Med
January 2025
Canadian Microbiology Proficiency Testing Program (CMPT), Department of Pathology and Laboratory Medicine, University of British Columbia, Vancouver, Canada.
External quality assessment (EQA) enhances patient safety through the evaluation of the quality of laboratory-based and point of care testing. Regulatory agencies and accreditation organizations utilize the results and the laboratory's response to them as part of assessing the laboratory's fitness to practice. In addition, where EQA samples are commutable and the assigned value has been determined using reference measurement procedures (RMPs), EQA data contributes to the verification of metrological traceability of assays as part of the post-market surveillance of diagnostic (IVD) medical devices (IVD-MDs).
View Article and Find Full Text PDFClin Chem Lab Med
January 2025
Canadian Microbiology Proficiency Testing Program (CMPT), Department of Pathology and Laboratory Medicine, University of British Columbia, Vancouver, Canada.
Providers of external quality assessment (EQA) programs evaluate data or information obtained and reported by participant laboratories using their routine procedures to examine properties or measurands in samples provided for this purpose. EQA samples must offer participants an equal chance to obtain accurate results, while being designed to provide results in clinically relevant ranges. It is the responsibility of the EQA provider to meet the necessary requirements for homogeneity, stability and some other properties of the EQA items in order to offer participants a fair, reliable and technically interesting EQA experience.
View Article and Find Full Text PDFPract Lab Med
November 2024
bioMérieux, Department of Research and Development for Immunoassays, Marcy-l'Étoile, France.
Introduction/objectives: Following the IFCC (The International Federation of Clinical Chemistry and Laboratory Medicine) guidelines concerning high-sensitivity cardiac troponin assays, we performed an assessment of the VIDAS® High-Sensitive Troponin I (TNHS) assay. The test was evaluated on its capacity to detect at least 50 % of healthy individuals and checked that the coefficient of variation was less than 10 % at the 99th percentile.
Methods: High-sensitivity performance was assessed by examining the limits of detection, the determination of the 99th percentile value, the evaluated imprecision at said value and the detectable results above limit of detection (LoD) in a cohort of healthy European individuals.