Background: The aim of this study was to compare body composition results from bioelectrical spectroscopy (BIS) with results from dual energy X-ray absorptiometry (DXA) in a population of male elite athletes. Body composition was assessed using DXA (Lunar Prodigy, GE Lunar Corp., Madison, USA) and BIS (Hydra 4200, Xitron Technologies Inc, San Diego, California, USA) at the same occasion. Agreement between methods was assessed using paired t-tests and agreement-plots.

Results: Thirty-three male elite athletes (soccer and ice hockey) were included in the study. The results showed that BIS underestimates the proportion of fat mass by 4.6% points in the ice hockey players. In soccer players the BIS resulted in a lower mean fat mass by 1.1% points. Agreement between the methods at the individual level was highly variable.

Conclusion: Body composition results assessed by BIS in elite athletes should be interpreted with caution, especially in individual subjects. BIS may present values of fat mass that is either higher or lower than fat mass assessed by DXA, independent of true fat content of the individual.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2267441PMC
http://dx.doi.org/10.1186/1477-5751-7-1DOI Listing

Publication Analysis

Top Keywords

body composition
16
elite athletes
16
fat mass
16
male elite
12
dual energy
8
energy x-ray
8
x-ray absorptiometry
8
composition assessed
8
assessed dxa
8
agreement methods
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!