Background: The quality of platelet (PLT) concentrates (PCs) can be evaluated with various in vitro methods. A new technique, free oscillation rheometry (FOR), can be used to monitor coagulation properties of PCs and gives information on clotting time and coagulum elasticity. This study compared the quality of apheresis PCs produced by the COBE Spectra and Trima Accel cell separators (both Gambro BCT) during storage for 7 days with in vitro tests including FOR.

Study Design And Methods: Apheresis PCs were collected with the COBE Spectra (n = 10) and Trima Accel (n = 10) cell separators. Swirling, blood gases, and metabolic variables were analyzed on Day 0. Samples taken on Days 1, 5, and 7 were also analyzed for hypotonic shock response (HSR), P-selectin, and glycoprotein (GP)Ib expression and evaluation of coagulation by FOR.

Results: Swirling, HSR, and percent GPIb-expressing PLTs were well maintained for 7 days, whereas glucose decreased and lactate increased significantly during storage for both Spectra and Trima PCs. Percent P-selectin-expressing cells increased to the same extent in both types of PCs during storage. pH increased between Day 0 and Day 1 but then decreased. The clotting time remained constant throughout the storage period whereas the development of elasticity was reduced on Days 5 and 7 compared to Day 1 (p < 0.05) for both types of PCs.

Conclusion: The results indicate that the PLT quality after storage for 7 days is well preserved, although activation of PLTs occurs during storage as assessed by in vitro tests. No difference in PLT quality was observed between Spectra- and Trima-produced PCs.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1537-2995.2007.01610.xDOI Listing

Publication Analysis

Top Keywords

spectra trima
16
cobe spectra
12
trima accel
12
accel cell
12
cell separators
12
storage days
12
quality platelet
8
produced cobe
8
assessed vitro
8
vitro methods
8

Similar Publications

Introduction: Collecting high-dose (HD) or double-dose (DD) apheresis platelets units from a single collection offers significant benefit by improving inventory logistics and minimizing the cost per unit produced. Platelet collection yield by apheresis is primarily influenced by donor factors, but the cell separator used also affects the collection yield.

Objectives: To predict the cutoff in donor factors resulting in HD and DD platelet collections between Trima/Spectra Optia and MCS+ apheresis equipment using Classification and Regression Trees (CART) analysis.

View Article and Find Full Text PDF

Background And Objectives: Platelet concentrates (PCs) can be prepared in several different ways, and they can be stored over few days before the use. Regulated on activation, normal T cells expressed and secreted (RANTES) levels in these concentrates may vary depending on the type of preparation and duration of storage of this component. We measured RANTES levels in platelet supernatants in different preparations and with different storage duration.

View Article and Find Full Text PDF

Background: Single donor apheresis platelets are superior in quality, but their usage is limited in a developing country due to cost and time constraints. Hence the product obtained must exceed in terms of yield, donor safety and technical convenience. Previous literature available on cell separators is on older versions.

View Article and Find Full Text PDF

Introduction: Referring to current standards the quality of an apheresis procedure is estimated by the quantity of collected cells. Nowadays a new kind of quality measurement could be found in the detection of cell volumina. Recent diagnostics have shown that stem cells and platelets - when separated - are likely to appear in a higher volume inside the cell product.

View Article and Find Full Text PDF

To evaluate the poststorage viability of apheresis platelets stored for up to 18 days in 80% platelet additive solution (PAS)/20% plasma, 117 healthy subjects donated platelets using the Haemonetics MCS+, COBE Spectra (Spectra), or Trima Accel (Trima) systems. Control platelets from the same subjects were compared with their stored test PAS platelets by radiolabeling their stored and control platelets with either (51)chromium or (111)indium. Trima platelets met Food and Drug Administration poststorage platelet viability criteria for only 7 days vs almost 13 days for Haemonetics platelets; ie, platelet recoveries after these storage times averaged 44 ± 3% vs 49 ± 3% and survivals were 5.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!