Purpose: The aim of this study was to determine the preload and tensile fracture load values of prosthetic retaining screws after long-term use in vivo compared to unused screws (controls). Additionally, the investigation addressed whether the preload and fracture load values of prosthetic retaining screws reported by the manufacturer become altered after long-term use in vivo.

Materials And Methods: For preload testing, 10 new screws (controls) from Nobel Biocare (NB) and 73 used retaining screws [58 from NB and 15 from Sterngold (SG)] were subjected to preload testing. For tensile testing, eight controls from NB and 58 used retaining screws (46 from NB and 12 from SG) were subjected to tensile testing. Used screws for both tests were in service for 18-120 months. A custom load frame, load cell, and torque wrench setup were used for preload testing. All 83 prosthetic screws were torqued once to 10 Ncm, and the produced preload value was recorded (N) using an X-Y plotter. Tensile testing was performed on a universal testing machine and the resulting tensile fracture load value was recorded (N). Preload and tensile fracture load values were analyzed with 2-way ANOVA and Tukey post-hoc tests.

Results: There was a significant difference between preload values for screws from NB and screws from SG (p < 0.001). The preload values for gold alloy screws from NB decreased as the number of years in service increased. There was a significant difference between tensile fracture values for the three groups (gold alloy screws from NB and SG and palladium alloy screws from NB) at p < 0.001. The tensile fracture values for gold alloy screws from NB and SG decreased as the number of years in service increased.

Conclusions: In fixed detachable hybrid prostheses, perhaps as a result of galling, the intended preload values of prosthetic retaining screws may decrease with increased in-service time. The reduction of the fracture load value may be related to the increase of in-service time; however, the actual determination of this relationship is not possible from this study alone.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1532-849X.2007.00290.xDOI Listing

Publication Analysis

Top Keywords

retaining screws
24
tensile fracture
24
fracture load
24
prosthetic retaining
16
screws
16
alloy screws
16
preload tensile
12
load values
12
values prosthetic
12
preload testing
12

Similar Publications

Load-bearing capacity of screw-retained fixed dental prostheses made of monolithic zirconia on different abutment designs and abutment-free implant connection.

J Dent

January 2025

Senior Research and Teaching Assistant, Clinic of Reconstructive Dentistry, Center for Dental Medicine, University of Zurich, Plattenstrasse 11, 8032 Zurich, Switzerland. Electronic address:

Objectives: A new abutment-free implant connection allows for direct screwing of FDPs on implants to avoid complications caused by cement rests or screw loosening, which may affect to screw torque and load distribution. The objective of this study was to test the initial (Fi) and final failure (Ff) loads and torque changes of abutment-free monolithic zirconia CAD-CAM fixed dental prostheses (FDPs) compared to titanium FDPs on different abutment designs.

Methods: Three-unit screw-retained FDPs (n=50) on two implants (n=100) were divided into groups (n=10) based on the implant-abutment connection and material of the supra-structure: (1) abutment-free monolithic CAD-CAM zirconia FDP (Abut-free-Zr), (2) abutment-free veneered titanium FDPs (Abut-free-Ti), (3) monolithic zirconia FDPs with titanium base abutments (Zr-Ti-Base), (4) monolithic zirconia FDPs on multi-unit abutments (Zr-MU), (5) veneered titanium FDP on multi-unit abutments (Ti-MU).

View Article and Find Full Text PDF

Purpose: This retrospective study aimed to compare extended sinus lift and extramaxilla surgical protocols for restoring severely atrophic maxillae with zygomatic implants (ZIs) and evaluate their clinical effectiveness.

Materials And Methods: The study includes patients who were treated at a dental clinic in Italy from 2012 to 2022. These patients received fixed screw-retained complete dentures supported by either two or four zygomatic implants (ZIs).

View Article and Find Full Text PDF

Objective: This study aimed to evaluate the clinical performance (degree of trueness) of a novel scan body "tooth-modified Scan body" (TMSB)& conventional scan body (CSB) in implant-supported full arch screw retained cases.

Methods: Seven edentulous arches (two maxillae, five mandibles) in 6 patients were rehabilitated with monolithic zirconia screw-retained implant prostheses supported by 4 (n = 1) and 5 implants (n = 6) for a total amount of 34 implants. Implant locations were scanned by intra-oral scanner (IOS) using two types of scan bodies, conventional scan bodies (CSB) in group (1) and tooth-modified scan bodies (TMSB) in group (2).

View Article and Find Full Text PDF

Immediate Implant Placement in a Hypothyroid Geriatric Patient: Aesthetic Excellence, Secondary Stability, and Successful Outcome-A Case Report.

Plast Aesthet Nurs (Phila)

December 2024

Sowmya Srinivas, BDS, MDS, PhD, is an Assistant Professor at the Department of Prosthodontics, JSS Dental College and Hospital, JSS Academy of Higher Education and Research, Mysore, Karnataka, India.

Placing an implant immediately into extraction sockets provides a distinct advantage over delayed placement, eliminating the need for a 4 to 6 months waiting period for bone formation. However, when patients present with hypothyroidism, the feasibility of immediate placement of dental implants becomes uncertain. This case involved a hypothyroid woman in her late 60s with loose lower anterior teeth.

View Article and Find Full Text PDF

Objective: To assess the clinical, radiographic and patient-reported outcome measures, and the success of screw-retained one-piece monolithic zirconia implant-supported restorations in the posterior region during a 1-year follow-up.

Methods: In a prospective case series, 50 single molar sites in the posterior region of 41 patients with a minimum age of 18 years and sufficient bone volume for placing an implant (≥ 8 mm) and space for an anatomical restoration were included. Following prosthetic-driven digital three-dimensional treatment planning, a tissue-level implant with an internal connection was inserted during a one-stage surgical procedure.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!