Objective: Two devices for self-measurement of blood pressure, one at the upper arm (Citizen CH-432B) and one at the wrist (Citizen CH-656C), were evaluated according to the International Protocol of the European Society of Hypertension.
Design: The International Protocol is divided into two phases: the first phase is performed on 15 selected participants with five participants in three different blood pressure ranges. If the devices passed this phase, 18 additional participants selected on the basis of the same criteria as in phase 1 were included.
Methods: Two skilled observers performed the following blood pressure measurements: five measurements were performed with the mercury standard alternating with four measurements with each of the test devices per participant. The first measurement result from each device instrument was not included in the analysis. The difference between the blood pressure value given by the devices and that obtained by the two observers (mean of the two observers) was calculated for each pair of measurements and classified into three categories (within 5, 10 and 15 mmHg). The results were compared to the pass criteria established by the European Society of Hypertension. Afterwards the number of measurement differences falling within 5 mmHg was determined for every person. At least 22 of the 33 participants should have two of their three comparisons within 5 mmHg and there should be a maximum of three participants without a measurement difference within the 5 mmHg range.
Results: Both tested devices passed the first phase of the validation process by exceeding the required number of comparisons falling within the 5, 10 and 15 mmHg error zones. Even the second phase confirmed the validation criteria with average differences between the device and the mercury sphygmomanometer of 0.7+/-4.4 and -3.6+/-4.0 mmHg for systolic blood pressure and diastolic blood pressure, respectively, for the Citizen CH-432B device and -0.7+/-6.0 and -1.2+/-4.5 mmHg for the Citizen CH-656C device. Phase 2 contains furthermore an individual analysis of the 33 participants, the requirements of which were also fulfilled by both devices.
Conclusion: The Citizen CH-432B and the Citizen CH-656C devices pass the validation recommendations of the International Protocol. They can be recommended for clinical use.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/MBP.0b013e3282cb599a | DOI Listing |
Circ Res
January 2025
Department of Physiology, Institute of Functional Genomics and Research Institute of Medical Science, Konkuk University School of Medicine, Chungju, Republic of Korea (H.L., S.P., J.R.A., M.S.S., H.J.N., B.K., Y.M.B.).
J Hypertens
December 2024
Department of Ultrasound Medicine, Tangdu Hospital, Air Force Medical University.
Background: The arterial stiffening is attributed to the intrinsic structural stiffening and/or load-dependent stiffening by increased blood pressure (BP). The respective lifetime alterations and major determinants of the two components with normal aging are not clear.
Methods: A total of 3053 healthy adults (1922 women) aged 18-79 years were enrolled.
J Hypertens
December 2024
University/British Heart Foundation Centre for Cardiovascular Science, The University of Edinburgh, Edinburgh, Scotland, UK.
Introduction: Hypertension is the leading preventable cause of cardiovascular morbidity and mortality globally, with a disproportionate impact on low-income and middle-income countries like Sri Lanka. Effective blood pressure (BP) control improves outcomes in patients with hypertension. This study aimed to assess the prevalence of uncontrolled hypertension, and its correlates among Sri Lankan patients with hypertension in clinic settings.
View Article and Find Full Text PDFJCEM Case Rep
January 2025
Division of Diabetology and Metabolism, Department of Internal Medicine, Tokyo Women's Medical University School of Medicine, Shinjuku-ku, Tokyo 162-8666, Japan.
A 37-year-old man presented with symptoms of polyuria and weight loss over the past year. Initial laboratory examination showed elevated blood glucose level (468 mg/dL [25.9 mmol/L]; normal reference range [RR], 75-109 mg/dL [4.
View Article and Find Full Text PDFJ Family Med Prim Care
December 2024
Department of Community Medicine, Smt. B.K. Shah Medical Institute and Research Centre, Sumandeep Vidyapeeth Deemed to be University, Piparia, Vadodara, Gujarat, India.
Introduction: Accredited Social Health Activists (ASHA) play a pivotal role as village-level functionaries under the National Health Mission, driving community transformation, facilitating health planning, and ensuring access to healthcare services.
Objectives: To assess the physical health status of ASHA workers and to investigate the prevalence of morbidities among ASHA workers.
Material And Methods: This is a cross-sectional study, conducted in Vadodara's east zone, and involves 123 ASHA workers.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!