Purpose: The sensitivity and specificity of enhanced photoemission spectroscopy (EPS) for performing an automated final check of compounded i.v. admixtures at a pediatric hospital pharmacy were studied.

Methods: A tabletop EPS device was used to test samples of seven high-risk drug-diluent combinations compounded in the pharmacy; the drugs were vancomycin, lorazepam, morphine, insulin, hydromorphone, gentamicin, and epinephrine. Ten sets of samples were prepared for each drug. Typically, a sample set consisted of dilutions ranging from 10-fold above to 10-fold below the targeted concentration. Testing was performed twice weekly between November 2005 and March 2006.

Results: The EPS device detected errors departing from the targeted concentration by 20% or more with a sensitivity of at least 95%. Specificity in distinguishing among test medications at targeted concentrations was 100%. The percentage of passing samples with intermediate concentrations varied among the drugs.

Conclusion: A tabletop EPS device demonstrated acceptable sensitivity and specificity for validating the identity and concentrations of selected high-risk i.v. medications compounded for pediatric patients. The device may help prevent clinically important medication errors caused by inaccurate compounding.

Download full-text PDF

Source
http://dx.doi.org/10.2146/ajhp060626DOI Listing

Publication Analysis

Top Keywords

eps device
12
enhanced photoemission
8
photoemission spectroscopy
8
high-risk medications
8
sensitivity specificity
8
tabletop eps
8
targeted concentration
8
spectroscopy verification
4
verification high-risk
4
medications purpose
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!