The use of a small precolumn instead of an injection loop for the determination of a new phytochemical drug, fellavine, and its metabolites is described. The method combines the direct injection of plasma and urine into the reversed-phase precolumn with separation on a Spheri-5 RP-18 analytical column. Different sorbents in the precolumn were compared. A recovery of fellavine and its metabolites from biological fluids except rat plasma of almost 100% was achieved on Chrompack RP (30-40 microns) and LiChrosorb RP-18 (7 microns). For rat plasma only the last sorbent gave 80% fellavine recovery. The influence of the protein binding on the fellavine recovery was examined. The limit of detection was equal to 0.05 micrograms/ml fellavine for plasma and 0.02 micrograms/ml for urine. To enhance the limit of detection longer precolumns were perferred.

Download full-text PDF

Source
http://dx.doi.org/10.1016/0378-4347(91)80461-kDOI Listing

Publication Analysis

Top Keywords

fellavine metabolites
12
rat plasma
12
determination phytochemical
8
phytochemical drug
8
drug fellavine
8
plasma urine
8
fellavine recovery
8
limit detection
8
fellavine
6
plasma
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!