The purpose of this single-center study was to report our initial experience with an implantable remote pressure sensor for aneurysm sac pressure measurement in patients post-endovascular aneurysm repair (EVAR) including short-term follow-up. A pressure sensor (EndoSure, Atlanta, GA) was implanted in 12 patients treated with different commercially available aortic endografts for EVAR. Pressure was read pre- and post-EVAR in the operating room. One-month follow-up (30 days+/-6 days) was performed including sac pressure readings and IV contrast CT scans. Variables were compared using the paired Student's t test. An intraprocedure type-I endoleak and a type-III endoleak were successfully treated resulting in decreasing sac pressures. In all patients, post-EVAR systolic sac pressure decreased by an average of 33% (P