Objective: To investigate the effect of acquisition arc (180 degrees or 360 degrees ) and attenuation correction (AC), in artifact occurrence and image uniformity of myocardial perfusion SPECT studies in normal hearts.

Methods: Twenty male and 44 female patients with primary hyperparathyroidism and low probability of resting perfusion abnormalities were analysed. Tc-sestamibi myocardial perfusion studies were performed at rest with a hybrid SPECT/CT system. An anthropomorphic phantom was also imaged.

Results: Without AC, a 360 degrees orbit resulted in significantly lower artifact rates compared to the standard 180 degrees in both gender groups. Differences persisted (P<0.001) even after adjusting reconstruction filter settings to compensate for disproportionate count statistics between the two arcs. However, gender-related variation was not considerably decreased and sub-diaphragmatic activity was augmented with a 360 degrees arc. After AC with either arc, substantially lower defect scores and minimal gender-related differences were obtained compared to the standard technique. A 360 degrees arc yielded lower defect rates and count variability than did the 180 degrees arc (P<0.05). These differences were small and were eliminated after weighting for dissimilar count statistics. Sub-diaphragmatic signal was amplified, particularly with 360 degrees ; severe bowel-myocardial activity overlap complicated image interpretation in 14% of cases with 360 degrees versus 8% with 180 degrees . In certain aspects, phantom imaging findings diverged from patient studies.

Conclusion: Without AC there is significant disparity between 180 degrees and 360 degrees concerning artifact occurrence. After CT-based AC differences between arcs in defect rate and image uniformity seem less critical. However, 180 degrees is favoured by less intensification of intestinal activity.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MNM.0b013e3282f28409DOI Listing

Publication Analysis

Top Keywords

180 degrees
12
360 degrees
12
degrees 360
8
myocardial perfusion
8
degrees
6
comparison 180
4
degrees acquisition
4
acquisition arcs
4
arcs correction
4
correction ct-based
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!