Objectives: To assess the long-term results of the Provox ActiValve, a prosthesis for voice rehabilitation after total laryngectomy that was designed to lessen the need for frequent replacements caused by Candida and "underpressure."

Patients And Methods: Retrospective assessment of device lifetime, indications for replacement, voice quality, and maintenance issues, measured by a structured trial specific questionnaire, in a cohort of 42 laryngectomized patients, experiencing a short Provox2 device lifetime (median, 21 days).

Results: The median device lifetime of Provox ActiValve, replaced for leakage through the device and those still in situ at the date of data collection (N = 32), was 337 days (mean 376 days): a statistically significant 16-fold increase compared to the Provox2 prosthesis (P < .001). In 10 patients, replacement was fistula-related (median after 86 days): esophageal pouch (N = 4), fistula granulation (N = 3), extrusion of the device (N = 2), and periprosthetic leakage (N = 1). Eighty-six percent of the patients used a special lubricant to diminish "stickiness" of the valve. Provox ActiValve was preferred by 90% of the patients who completed the trial-specific questionnaire.

Conclusions: For patients requiring frequent device-related replacements, Provox ActiValve, also long-term, provides a true solution and thereby is a valuable addition to prosthetic voice rehabilitation.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MLG.0b013e318159ebdeDOI Listing

Publication Analysis

Top Keywords

provox activalve
20
device lifetime
12
long-term provox
8
voice rehabilitation
8
activalve
5
device
5
patients
5
activalve solving
4
solving problem
4
problem frequent
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!