Background: More frequent haemodialysis can improve both survival and quality of life of patients with chronic kidney disease. However, there is little capacity in the UK to allow patients to have more frequent haemodialysis treatments in hospital and satellite haemodialysis units. New means of delivering haemodialysis are therefore required. Our aim was to assess the safety and efficiency of a wearable haemodialysis device.

Methods: Eight patients with end-stage kidney failure (five men, three women, mean age 51.7 [SD 13.8] years) who were established on regular haemodialysis were fitted with a wearable haemodialysis device for 4-8 h. Patients were given unfractionated heparin for anticoagulation, as they would be for standard haemodialysis.

Findings: There were no important cardiovascular changes and no adverse changes in serum electrolytes or acid-base balance. There was no evidence of clinically significant haemolysis in any patient. Mean blood flow was 58.6 (SD 11.7) mL/min, with a dialysate flow of 47.1 (7.8) mL/min. The mean plasma urea clearance rate was 22.7 (5.2) mL/min and the mean plasma creatinine clearance rate was 20.7 (4.8) mL/min. Clotting of the vascular access occurred in two patients when the dose of heparin was decreased and the partial thromboplastin time returned towards the normal reference range in both of these patients. The fistula needle became dislodged in one patient, but safety mechanisms prevented blood loss, the needle was replaced, and treatment continued.

Interpretation: This wearable haemodialysis device shows promising safety and efficacy results, although further studies will be necessary to confirm these results.

Download full-text PDF

Source
http://dx.doi.org/10.1016/S0140-6736(07)61864-9DOI Listing

Publication Analysis

Top Keywords

wearable haemodialysis
16
haemodialysis device
12
patients end-stage
8
haemodialysis
8
frequent haemodialysis
8
ml/min plasma
8
clearance rate
8
patients
7
wearable
4
device patients
4

Similar Publications

Anaemia is a prevalent complication in patients with end-stage kidney disease (ESKD) undergoing haemodialysis. This study evaluates the accuracy of the Alio SmartPatch™, a non-invasive remote monitoring device, in measuring haemoglobin (Hb) and haematocrit (Hct) levels in haemodialysis patients by comparing its results with standard blood-based laboratory methods. The results from 116 patients across multiple sites in the USA and the Kingdom of Jordan show that SmartPatch measurements align closely with standard blood-based laboratory methods, meeting clinically acceptable limits of agreement.

View Article and Find Full Text PDF

This study aimed to develop a real-time, noninvasive hyperkalemia monitoring system for dialysis patients with chronic kidney disease. Hyperkalemia, common in dialysis patients, can lead to life-threatening arrhythmias or sudden death if untreated. Therefore, real-time monitoring of hyperkalemia in this population is crucial.

View Article and Find Full Text PDF

Background: Accurate assessment of fluid volume and hydration status is essential in many disease states, including patients with chronic kidney disease. The aim of this study was to investigate the ability of a wearable continuous bioimpedance sensor to detect changes in fluid volume in patients undergoing regular hemodialysis (HD).

Methods: 31 patients with end-stage renal disease were enrolled and monitored with a sensor patch (Re:Balans) on the upper back through two consecutive HD sessions and the interdialytic period between.

View Article and Find Full Text PDF

A novel LC-MS/MS assay for low concentrations of creatinine in sweat and saliva to validate biosensors for continuous monitoring of renal function.

J Chromatogr B Analyt Technol Biomed Life Sci

December 2024

Clinical Laboratory, Catharina Hospital Eindhoven, Eindhoven 5623 EJ, The Netherlands; Department of Biomedical Engineering, Chemical Biology, Eindhoven University of Technology, Groene Loper 3, Eindhoven 5612 AE, The Netherlands.

Monitoring of kidney function traditionally relies on plasma creatinine concentrations, necessitating invasive blood draws. Non-invasively obtainable biofluids, such as sweat and saliva, present a patient-friendly alternative with potential for continuous monitoring. This study focusses on developing and validating a novel Liquid Chromatography- tandem Mass Spectrometry (LC-MS/MS) assay as a reference test for measuring low creatinine concentrations in sweat and saliva.

View Article and Find Full Text PDF

Hemodialysis for chronic kidney disease (CKD) relies on vascular access (VA) devices, such as arteriovenous fistulas (AVF), grafts (AVG), or catheters, to maintain blood flow. Nonetheless, unpredictable progressive vascular stenosis due to neointimal formation or complete occlusion from acute thrombosis remains the primary cause of mature VA failure. Despite emergent surgical intervention efforts, the lack of a reliable early detection tool significantly reduces patient outcomes and survival rates.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!