Objectives: This study sought to validate a new method for direct volumetric blood flow measurement in coronary arteries in animals and in conscious humans during cardiac catheterization.

Background: Direct volumetric measurement of blood flow in selective coronary arteries would be useful for studying the coronary circulation.

Methods: Based on the principle of thermodilution with continuous low-rate infusion of saline at room temperature, we designed an instrumental setup for direct flow measurement during cardiac catheterization. A 2.8-F infusion catheter and a standard 0.014-inch sensor-tipped pressure/temperature guidewire were used to calculate absolute flow (Q(thermo)) in a coronary artery from the infusion rate of saline, temperature of the saline at the tip of the infusion catheter, and distal blood temperature during infusion. The method was tested over a wide range of flow rates in 5 chronically instrumented dogs and in 35 patients referred for physiological assessment of a coronary stenosis or for percutaneous coronary intervention.

Results: Thermodilution-derived flow corresponded well with true flow (Q) in all dogs (Q(thermo) = 0.73 Q + 42 ml/min; R(2) = 0.72). Reproducibility was excellent (Q(thermo,)(1) = 0.96 x Q(thermo,)(2) + 3 ml/min; R(2) = 0.89). The measurements were independent of infusion rate and sensor position as predicted by theory. In the humans, a good agreement was found between increase of thermodilution-derived volumetric blood flow after percutaneous coronary intervention and increase of fractional flow reserve (R(2) = 0.84); reproducibility of the measurements was excellent (Q(thermo,)(1) = 1.0 Q(thermo,)(2) + 0.9 ml/min, R(2) = 0.97), and the measurements were independent of infusion rate and sensor position.

Conclusions: Using a suitable infusion catheter and a 0.014-inch sensor-tipped guidewire for measurement of coronary pressure and temperature, volumetric blood flow can be directly measured in selective coronary arteries during cardiac catheterization.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jacc.2007.08.047DOI Listing

Publication Analysis

Top Keywords

blood flow
20
volumetric blood
16
coronary arteries
16
direct volumetric
12
flow measurement
12
measurement coronary
12
infusion catheter
12
infusion rate
12
flow
11
coronary
10

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!