Introduction: During commissioning of the Pinnacle (Philips) treatment planning system (TPS) the margining algorithm was investigated and was found to produce larger PTVs than Plato (Nucletron) for identical GTVs. Subsequent comparison of PTV volumes resulting from the QA outlining exercise for the CHHIP (Conventional or Hypofractionated High Dose IMRT for Prostate Ca.) trial confirmed that there were differences in TPS's margining algorithms. Margining and the clinical impact of the different PTVs in seven different planning and virtual simulation systems (Pinnacle, Plato, Prosoma (MedCom), Eclipse (7.3 and 7.5) (Varian), MasterPlan (Nucletron), Xio (CMS) and Advantage Windows (AW) (GE)) is investigated, and a simple test for 3D margining consistency is proposed.

Methods: Using each TPS, two different sets of prostate GTVs on 2.5mm and 5mm slices were margined according to the CHHIP protocol to produce PTV3 (prostate+5 mm/0 mm post), PTV2 (PTV3+5 mm) and PTV1 (prostate and seminal vesicles+10 mm). GTVs and PTVs were imported into Pinnacle for volume calculation. DVHs for 5mm slice plans, created using the smallest PTVs, were recalculated on the largest PTV dataset and vice versa. Since adding a margin of 50 mm to a structure should give the same result as adding five margins of 10 mm, this was tested for each TPS (consistency test) using an octahedron as the GTV and CT datasets with 2.5 mm and 5 mm slices.

Results: The CHHIP PTV3 and PTV1 volumes had a standard deviation, across the seven systems, of 5% and PTV2 (margined twice) 9%, on the 5 mm slices. For 2.5 mm slices the standard deviations were 4% and 6%. The ratio of the Pinnacle and the Eclipse 7.3 PTV2 volumes was 1.25. Rectal doses were significantly increased when encompassing Pinnacle PTVs (V(50)=42.8%), compared to Eclipse 7.3 PTVs (V(50)=36.4%). Conversely, fields that adequately treated an Eclipse 7.3 PTV2 were inadequate for a Pinnacle PTV2. AW and Plato PTV volumes were the most consistent (0.3%) and (-0.4%). However, the 1x50mm margin in Pinnacle produced a 15.9% larger volume than 5 x 10 mm margins, while for Eclipse 7.3 the single margined volume was 14.3% smaller. These inconsistencies were reduced to approximately 5% by adjusting the superior/inferior margins.

Conclusions: Accurate margin algorithms are necessary to ensure that volume expansion does not add extra uncertainty to the radiotherapy planning process. We have found significant differences in the 3D margining algorithms of TPSs, devised a simple test to predict inconsistency and suggested corrective action to minimise the variation.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.radonc.2007.11.006DOI Listing

Publication Analysis

Top Keywords

margining algorithms
12
treatment planning
8
ptv volumes
8
simple test
8
eclipse ptv2
8
pinnacle
7
ptvs
6
margining
5
eclipse
5
ptv2
5

Similar Publications

A multiple model approach to respiratory motion prediction for real-time IGRT.

Phys Med Biol

March 2008

Control Theory and Applications Centre, Coventry University, Priory Street, Coventry CV1 5FB, UK.

Respiration induces significant movement of tumours in the vicinity of thoracic and abdominal structures. Real-time image-guided radiotherapy (IGRT) aims to adapt radiation delivery to tumour motion during irradiation. One of the main problems for achieving this objective is the presence of time lag between the acquisition of tumour position and the radiation delivery.

View Article and Find Full Text PDF

Introduction: During commissioning of the Pinnacle (Philips) treatment planning system (TPS) the margining algorithm was investigated and was found to produce larger PTVs than Plato (Nucletron) for identical GTVs. Subsequent comparison of PTV volumes resulting from the QA outlining exercise for the CHHIP (Conventional or Hypofractionated High Dose IMRT for Prostate Ca.) trial confirmed that there were differences in TPS's margining algorithms.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!