A new procedure for automatic separation and preconcentration of 241Am and 239+240Pu from interfering matrixes using transuranide (TRU)-resin is proposed. Combination of the multisyringe flow injection analysis and multipumping flow system techniques with the TRU-resin allows carrying out the sampling treatment and separation in a short time using large sample volumes. Americium is eluted from the column with 4 mol L(-1) hydrochloric acid, and then plutonium is separated via on-column Pu(IV) reduction to Pu(III) with titanium(III) chloride. The corresponding alpha activities are measured off-line, with a relative standard deviation of 3% and a lower limit of detection of 0.004 Bq mL(-1), by using a multiplanchet low-background proportional counter.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1021/ac070725m | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!