Objective: To evaluate the interindividual and intraindividual applicator position variability in high dose rate ring and tandem intracavitary brachytherapy applications in locally advanced cervical cancer.

Patients And Methods: Eight patients with locally advanced cervical cancer formed the study population. They had been treated in Dokuz Eylül University Department of Radiation Oncology between the years 2000 and 2005 with high dose rate intracavitary brachytherapy using ring and tandem applicators. The 3-dimensional geometric variation of the applicator center in craniocaudal, mediolateral and anteroposterior directions was determined on the basis of bony reference points in 24 pairs of orthogonal films obtained in the conventional simulator. Then the following evaluations were performed: 1) the applicator position variability in all applications (interindividual variability), 2) the intraindividual applicator position variability relative to the first application, 3) the intraindividual applicator position variability relative to the average of three applications. Among the potential factors that might influence the reproducibility of ring and tandem applications, age, stage, the period between external radiotherapy and brachytherapy were evaluated by univariate analysis.

Results: Standard deviation of interindividual applicator variability was 3.83 mm in craniocaudal, 0.39 mm in mediolateral and 2.86 mm in anteroposterior directions. The standard deviation of intraindividual variability relative to the first application was 1.91 mm in craniocaudal, 0.4 mm in mediolateral, and 4.26 mm in anteroposterior directions. The standard deviation of intraindividual variability relative to the average of three applications was 0.95 mm in craniocaudal, 1.86 mm in mediolateral, and 1.24 mm in anteroposterior directions. According to univariate analysis, no factor influenced applicator position variability.

Conclusions: In order to extract definitive conclusions about factors that affect positional reproducibility of ring and tandem applicators, studies are needed that include larger numbers of patients.

Download full-text PDF

Source
http://dx.doi.org/10.1177/030089160709300505DOI Listing

Publication Analysis

Top Keywords

applicator position
24
ring tandem
20
position variability
16
anteroposterior directions
16
variability relative
16
high dose
12
dose rate
12
intracavitary brachytherapy
12
tandem applicators
12
intraindividual applicator
12

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!