Coronary calcification: effect of small variation of scan starting position on Agatston, volume, and mass scores.

Radiology

Department of Radiology, University Medical Center Utrecht, Heidelberglaan 100, Room E01.132, 3584 CX Utrecht, The Netherlands.

Published: January 2008

Purpose: To retrospectively evaluate the effect of a small variation of scan starting position on coronary artery calcium scores based on nonoverlapping 3-mm multidetector computed tomographic (CT) data sets.

Materials And Methods: Informed consent and institutional review board approval were obtained. A retrospective study was performed by using prospective unenhanced electrocardiographically triggered cardiac multidetector CT scans in 228 women (mean age, 67 years +/- 5 [standard deviation]). From the original 1.5-mm data set, two sets of adjacent images with a section thickness of 3 mm and a variation in starting point of 1.5 mm were obtained. Calcium scoring was performed to acquire Agatston, volume, and mass scores. Subjects were assigned to one of five risk categories (I-V) according to the Agatston score of each 3-mm data set and the average score. Kappa value was calculated to assess agreement in risk category assignment. Differences and relative differences between scores obtained for both 3-mm data sets were calculated overall and according to risk category. The effect of scoring algorithm on the relative differences between scores was analyzed with the Wilcoxon signed rank test.

Results: Categories I-V contained 102, 35, 48, 31, and 12 subjects, respectively. For all scoring algorithms, median relative differences decreased from more than 130% in category II to less than 10% in category V. In the three highest categories, relative differences were significantly smaller for volume and mass scores than for Agatston scores (P < .05). Twenty-one subjects were assigned to different risk categories between the two data sets (kappa = 0.87). Eleven patients were assigned a nonzero score in one and a zero score in the other data set.

Conclusion: A small variation in scan starting position can substantially influence calcium measurements and poses an inherent limit to calcium scoring with contiguous 3-mm sections. Mass and volume scores are slightly less affected than are Agatston scores.

Download full-text PDF

Source
http://dx.doi.org/10.1148/radiol.2461070006DOI Listing

Publication Analysis

Top Keywords

relative differences
16
small variation
12
variation scan
12
scan starting
12
starting position
12
volume mass
12
mass scores
12
scores
9
agatston volume
8
data set
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!