Background: Occipital pressure ulcers are well-known complications for trauma patients wearing cervical collars. We assessed the effects of four commercial cervical collars (Aspen, Philadelphia, Miami J, and Miami J with Occian back [Miami J/Occian]) on cervical range of motion (CROM) and mandibular and occipital tissue-interface pressure (TIP).
Methods: Forty-eight healthy volunteers (24 men, 24 women; mean age, 38.9 years +/- 10.5 years) were stratified by body mass index. CROM was measured in the seated position without and with collars. Sagittal, coronal, and rotatory CROM was measured with a goniometer. Occipital and mandibular pressures were mapped with subjects in upright and supine positions.
Results: All collars significantly restricted CROM in all planes (p < 0.001). The Philadelphia and standard Miami J collars were the most restrictive. The Aspen collar was the least restrictive for flexion and rotation. The Miami J/Occian back was the least restrictive for extension and lateral flexion. For supine measurements, Miami J and Miami J/Occian back had the lowest mean TIP, whereas Aspen and Philadelphia collars had the greatest (p < 0.001). For upright measurements, the Miami J/Occian back produced the smallest mean TIPs; the other collars, ranked by ascending TIP, were Philadelphia, Miami J, and Aspen (p < 0.001). Philadelphia and Miami J collars had significant collar-body mass index interaction effects on supine occiput mean pressure (p = 0.04).
Conclusions: Miami J and Philadelphia collars restricted CROM to the greatest extent. Miami J and Miami J/Occian back had the lowest levels of mandibular and occipital pressure; these collars may markedly reduce the risk of occipital pressure ulcers without compromising immobilization.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/TA.0b013e3180487d0f | DOI Listing |
J Trauma
November 2007
Department of Nursing, Mayo Clinic, Rochester, Minnesota 55905, USA.
Background: Occipital pressure ulcers are well-known complications for trauma patients wearing cervical collars. We assessed the effects of four commercial cervical collars (Aspen, Philadelphia, Miami J, and Miami J with Occian back [Miami J/Occian]) on cervical range of motion (CROM) and mandibular and occipital tissue-interface pressure (TIP).
Methods: Forty-eight healthy volunteers (24 men, 24 women; mean age, 38.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!