A global calibration model for a-Si EPIDs used for transit dosimetry.

Med Phys

Department of Radiation Oncology (MAASTRO), Research Institute for Growth and Development, University Hospital of Maastricht, Maastricht, The Netherlands.

Published: October 2007

Electronic portal imaging devices (EPIDs) are not only applied for patient setup verification and detection of organ motion but are also increasingly used for dosimetric verification. The aim of our work is to obtain accurate dose distributions from a commercially available amorphous silicon (a-Si) EPID for transit dosimetry applications. For that purpose, a global calibration model was developed, which includes a correction procedure for ghosting effects, field size dependence and energy dependence of the a-Si EPID response. In addition, the long-term stability and additional buildup material for this type of EPID were determined. Differences in EPID response due to photon energy spectrum changes have been measured for different absorber thicknesses and field sizes, yielding off-axis spectrum correction factors based on transmission measurements. Dose measurements performed with an ionization chamber in a water tank were used as reference data, and the accuracy of the dosimetric calibration model was determined for a large range of treatment conditions. Gamma values using 3% as dose-difference criterion and 3 mm as distance-to-agreement criterion were used for evaluation. The field size dependence of the response could be corrected by a single kernel, fulfilling the gamma evaluation criteria in case of virtual wedges and intensity modulated radiation therapy fields. Differences in energy spectrum response amounted up to 30%-40%, but could be reduced to less than 3% using our correction model. For different treatment fields and (in)homogeneous phantoms, transit dose distributions satisfied in almost all situations the gamma criteria. We have shown that a-Si EPIDs can be accurately calibrated for transit dosimetry purposes.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.2776244DOI Listing

Publication Analysis

Top Keywords

calibration model
12
transit dosimetry
12
global calibration
8
a-si epids
8
dose distributions
8
a-si epid
8
field size
8
size dependence
8
epid response
8
energy spectrum
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!