Background And Aim Of The Study: The new Sorin Freedom SOLO pericardial stentless valve is designed for supra-annular implantation, and requires only one running suture. It can be implanted with a short cross-clamp time, and is designed to offer the same hemodynamic advantages of other stentless valves. The study aim was to evaluate the prospective postoperative and two-month follow up hemodynamic performance of this bioprosthesis.
Methods: Thirty patients (13 males, 17 females; mean age 75.6 +/- 6.21 years) with severe aortic stenosis underwent valve replacement with the Sorin Freedom SOLO stentless valve. All patients underwent transthoracic echocardiography before surgery, before hospital discharge, and at two months' follow up. The peak and mean transprosthetic gradients, telediastolic and telesystolic diameters, septal and posterior wall thicknesses, total and indexed ventricular mass volume and left ventricular ejection fraction were evaluated.
Results: Both, the transprosthetic peak gradient and mean gradient decreased significantly during the first two months (p < 0.05 and p < 0.001, respectively). The telediastolic diameter was significantly reduced between preoperative evaluation and follow up (p < 0.05). The interventricular septum thickness was decreased significantly after two months (p <0.001), as was the posterior wall thickness, albeit to a lesser degree (p < 0.05). Both, total and indexed ventricular mass volume showed a significant regression at the two months follow up (p < 0.001).
Conclusion: The Sorin Freedom SOLO stentless valve shows good hemodynamic performance, with an early and highly progressive left ventricular remod eling. If these data are confirmed in future studies, the SOLO prosthesis might represent a safe alternative to the use of conventional stentless valves.
Download full-text PDF |
Source |
---|
Am J Case Rep
July 2024
Department of Morphology and Genetics, Escola Paulista de Medicina, Universidade Federal de São Paulo (UNIFESP), São Paulo, SP, Brazil.
BACKGROUND A blood cell saver, or autotransfusion system, is used to collect, wash, and return autologous blood collected from the surgical patient. This report describes a 55-year-old man who underwent combined mitral and aortic valve replacement surgery with cardiopulmonary bypass and had a successful outcome following intraoperative and postoperative autologous blood transfusion using a blood cell saver. CASE REPORT The patient did not accept blood transfusion for reasons of religious conscience and was in a critical condition, receiving palliative care.
View Article and Find Full Text PDFCardiovasc Pathol
June 2021
Department of Cardiac, Thoracic and Vascular Sciences and Public Health, University of Padua, Italy. Electronic address:
Background: Aortic stentless bioprosthetic valve (SLBPV), either porcine or pericardial, minimizes transvalvular gradient and favors regression of left ventricular hypertrophy. The drawback consists of longer time for suturing. While structural valve deterioration (SVD) in stented porcine and pericardial BPVs has been extensively investigated, less information is available on SLBPVs.
View Article and Find Full Text PDFJ Thorac Cardiovasc Surg
February 2021
Children's Hospital of Philadelphia, Philadelphia, Pa.
Objectives: To assess the difference in time to and predictors of reintervention according to valve type in surgical bioprosthetic pulmonary valve replacement (PVR) in patients younger than 30 years of age from multiple centers.
Methods: Data were retrospectively collected for 1278 patients <30 years of age undergoing PVR at 8 centers between 1996 and 2015.
Results: Mean age at PVR was 19.
Innovations (Phila)
November 2021
47291 Department of Cardiac Surgery, European Hospital, Rome, Italy.
Heart Lung Circ
March 2021
Clinic for Cardiovascular Surgery, Inselspital, University Hospital Bern, Bern, Switzerland.
Objectives: Postoperative permanent pacemaker implantation (PPI) after conventional aortic valve replacement (AVR), due to new-onset severe conduction system disorders, is required in approximately 7% of patients. This study investigated the need for PPI after AVR with conventional stented Perimount Magna compared with the Freedom SOLO (FS) stentless valve (Sorin Group, Saluggia, Italy), now LivaNova plc (London, UK) that uses a strictly supra-annular, subcoronary running suture implantation technique, sparing the vulnerable interleaflet triangles in the region of the septum membranaceum.
Methods: A total of 413 consecutive patients (71.