Gaseous emboli detection based on a dual-wavelet transform analysis.

Proc Inst Mech Eng H

Department of Biomedical Engineering, Engineering College of Aarhus, Aarhus, Denmark.

Published: August 2007

Emboli monitoring is nowadays based on the assessment of microembolic signals by Doppler ultrasound. However, the present systems have problems in detecting multiple emboli. A more dedicated algorithm for post-processing of the recorded Doppler signals was proposed. Based on the hypothesis that single and multiple gaseous emboli can be quantified by combining discrete and continuous wavelet transformation, the aim of this study was to detect gaseous emboli and to validate our method visually. A flow rig was used where gaseous emboli were generated. Doppler signals and visual validation data of gaseous emboli were acquired simultaneously. Microembolic signals were extracted and analysed using wavelet transformation. Results were validated against a visual reference. At various degrees of bubble generation, the system had 100 per cent detection during a low frequency of bubble generation but an estimation error of 7.4 per cent during a high frequency of bubble generation. The estimation error varied between -7.4 and +3 per cent. The system had a higher rate of success in detecting large gaseous emboli in small numbers than small gaseous emboli in large numbers. Single and double emboli were successfully detected and separated, whereas gaseous emboli clouds could be detected but not quantified. Being able to separate simultaneous gaseous emboli may offer new means of increasing detectability for embolism monitoring.

Download full-text PDF

Source
http://dx.doi.org/10.1243/09544119JEIM295DOI Listing

Publication Analysis

Top Keywords

gaseous emboli
36
bubble generation
12
emboli
11
gaseous
9
microembolic signals
8
doppler signals
8
wavelet transformation
8
frequency bubble
8
generation estimation
8
estimation error
8

Similar Publications

Article Synopsis
  • * Results showed that NO significantly reduced both the number and volume of microemboli compared to a control group, with a decrease from 55,478 to 1,197 microemboli over one hour (p=0.016).
  • * Additionally, the group receiving NO exhibited lower levels of neuron-specific enolase, indicating reduced brain injury, with concentrations dropping to 7.7 ng/ml, compared to 11.2 ng/ml in the control group (p=0.047).
View Article and Find Full Text PDF

Aims: The significance of micro-embolic signals (MESs) during atrial fibrillation (AF) ablation is unclear. Previous studies had limitations, and cryoballoon (CB) ablation patients were under-represented. Minimizing MESs is recommended due to their uncertain neurocognitive impact.

View Article and Find Full Text PDF

Hydrogen peroxide (H₂O₂) ingestion can lead to severe systemic complications, including neurological sequelae such as acute embolic stroke. We present a case of a 49-year-old male who accidentally ingested approximately 50-60 mL of 50% w/w hydrogen peroxide, resulting in encephalopathy, upper motor neuron quadriparesis, and pulmonary artery thrombosis. The patient's altered sensorium progressed to a stupor, accompanied by acute respiratory distress and abdominal gaseous distension.

View Article and Find Full Text PDF

Xylem cell size regulation is a key adaptive response to water deficit in Eucalyptus grandis.

Tree Physiol

July 2024

Institute for Plant Biotechnology, Department of Genetics, Stellenbosch University, Private Bag X1, Matieland, Stellenbosch 7602, South Africa.

Future climatic scenarios forecast increasingly frequent droughts that will pose substantial consequences on tree mortality. In light of this, drought-tolerant eucalypts have been propagated; however, the severity of these conditions will invoke adaptive responses, impacting the commercially valuable wood properties. To determine what mechanisms govern the wood anatomical adaptive response, highly controlled drought experiments were conducted in Eucalyptus grandis W.

View Article and Find Full Text PDF

Background: During cardiac surgery the use of a minimal extracorporeal circulation (MiECC) system may reduce the adverse effects for the patient. This is probably caused by reduced inflammation and hemodilution. For the use of a MiECC circuit, a venous bubble trap (VBT) is warranted for safety reasons.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!