Objective: To describe and report the reliability of a portable, laptop-based, real-time, continuous physiologic data acquisition system (PDAS) that allows for synchronous recording of physiologic data, clinical events, and event markers at the bedside for physiologic research studies in the intensive care unit.

Design: Descriptive report of new research technology.

Setting: Adult and pediatric intensive care units in three tertiary care academic hospitals.

Patients: Sixty-four critically ill and injured patients were studied, including 34 adult (22 males and 12 females) and 30 pediatric (19 males and 11 females).

Interventions: None.

Measurements And Main Results: Data transmission errors during bench and field testing were measured. The PDAS was used in three separate research studies, by multiple users, and for repeated recordings of the same set of signals at various intervals for different lengths of time. Both parametric (1 Hz) and waveform (125-500 Hz) signals were recorded and analyzed. Details of the PDAS components are explained and examples are given from the three experimental physiology-based protocols. Waveform data include electrocardiogram, respiration, systemic arterial pressure (invasive and noninvasive), oxygen saturation, central venous pressure, pulmonary arterial pressure, left and right atrial pressures, intracranial pressure, and regional cerebral blood flow. Bench and field testing of the PDAS demonstrated excellent reliability with 100% accuracy and no data transmission errors. The key feature of simultaneously capturing physiologic signal data and clinical events (e.g., changes in mechanical ventilation, drug administration, clinical condition) is emphasized.

Conclusions: The PDAS provides a reliable tool to record physiologic signals and associated clinical events on a second-to-second basis and may serve as an important adjunctive research tool in designing and performing clinical physiologic studies in critical illness and injury.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.PCC.0000288715.66726.64DOI Listing

Publication Analysis

Top Keywords

physiologic data
12
clinical events
12
data acquisition
8
acquisition system
8
data clinical
8
physiologic studies
8
intensive care
8
data transmission
8
transmission errors
8
bench field
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!