Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Automated collection of blood components offers multiple advantages and has prompted development of portable devices. This study sought to document the biochemical and hematologic properties and in vivo recovery of red cells (RBCs) collected via a new device that employed a variable-volume centrifugal separation chamber.
Study Design And Methods: Normal subjects (n = 153) donated 2 units of RBCs via an automated blood collection system (Cymbal, Haemonetics). Procedures were conducted with wall outlet power (n = 49) or the device's battery source (n = 104). Units were collected with or without leukoreduction filtration and were stored in AS-3 for 42 days. The units were assessed via standard biochemical and hematologic tests before and after storage, and 24 leukoreduced (LR) and 24 non-LR RBCs were radiolabeled on Day 42 with Na(2)(51)CrO(4) for autologous return to determine recovery at 24 hours with concomitant determination of RBC volume via infusion of (99m)Tc-labeled fresh RBCs.
Results: Two standard RBC units (targeted to contain 180 mL of RBCs plus 100 mL of AS-3) could be collected in 35.7 +/- 2.0 minutes (n = 30) or 40.3 +/- 2.7 minutes for LR RBCs (n = 92). An additional 31 collections were conducted successfully with intentional filter bypassing. RBC units contained 104 +/- 4.1 percent of their targeted volumes (170-204 mL of RBCs), and LR RBCs contained 92 percent of non-LR RBCs' hemoglobin. All LR RBCs contained less than 1 x 10(6) white blood cells. Mean hemolysis was below 0.8 percent (Day 42) for all configurations. Adenosine triphosphate was well preserved. Mean recovery was 82 +/- 4.9 percent for RBCs and 84 +/- 7.0 percent for LR RBCs.
Conclusions: The Cymbal device provided quick and efficient collection of 2 RBC units with properties meeting regulatory requirements and consistent with good clinical utility.