Purpose: To determine whether the HoMedics Model BPA-300 meets American Association for the Advancement of Medical Instrumentation (AAMI) accuracy standards.

Method: Simultaneous auscultatory measurements by two observers using a mercury manometer were obtained in participants during measurement with the HoMedics Model BPA-300. Triple measurements were obtained after a minimum of 5 min of rest and 1 min between readings.

Results: In total, 132 participants had 396 nurse-measured blood pressures (BP) and 385 successful systolic and 383 diastolic monitor readings. The initial auscultation systolic BP was less than 100 mmHg in 17 participants (12.9%), between 100 and 160 mmHg in 108 participants (81.8%), and greater than 160 mmHg in seven (5.3%) of participants. The average difference between the observers and device was 0.6+/-4.3 mmHg for systolic and 0.5+/-4.3 mmHg for diastolic-measured BPs by AAMI method 1, which met this guideline. The average difference between observers and device by AAMI method 2 was 0.9+/-4.0 mmHg for systolic and 0.5+/-3.9 mmHg for diastolic-measured BP, which met this guideline.

Conclusion: The HoMedics Model BPA-300 meets AAMI accuracy standards using methods 1 and 2.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MBP.0b013e3282c9ac12DOI Listing

Publication Analysis

Top Keywords

homedics model
12
model bpa-300
12
bpa-300 meets
8
aami accuracy
8
160 mmhg
8
average difference
8
difference observers
8
observers device
8
mmhg systolic
8
mmhg diastolic-measured
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!