[Monitor tables for electron beams in radiotherapy].

Z Med Phys

Klinik für Radioonkologie, Universitätsklinikum Tübingen, Medizinische Physik.

Published: October 2007

The application of electron beams in radiotherapy is still based on tables of monitor units, although 3-D treatment planning systems for electron beams are available. This have several reasons: The need for 3-D treatment planning is not recognized; there is no confidence in the calculation algorithm; Monte-Carlo algorithms are too time-consuming; and the effort necessary to measure basic beam data for 3-D planning is considered disproportionate. However, the increasing clinical need for higher dosimetric precision and for more conformal electron beams leads to the requirement for more sophisticated tables of monitor units. The present paper summarizes and discusses the main aspects concerning the preparation of tables of monitor units for electron beams. The measurement equipment and procedures for measuring basic beam data needed for tables of monitor units for electron beams are described for a standard radiation therapy linac. The design of tables of monitor units for standard electron applicators is presented; this design can be extended for individual electron inserts, to variable applicator surface distances, to oblique beam incidence, and the use of bolus material. Typical data of an Elekta linac are presented in various tables.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.zemedi.2007.03.004DOI Listing

Publication Analysis

Top Keywords

electron beams
24
tables monitor
20
monitor units
20
electron
8
3-d treatment
8
treatment planning
8
basic beam
8
beam data
8
units electron
8
beams
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!