Background: Cardiac resynchronization therapy (CRT) is an established method for treatment of patients with severe congestive heart failure and asynchronous left ventricular contraction. Its clinical and haemodynamic benefits are well documented. We have retrospectively reviewed CRT performed at our centre.

Material And Methods: Our department treated 150 patients with CRT according to accepted indications from 1999-2006. The patients were observed for two years and one third reached two-year follow-up. We reviewed operative complications, lead re-implantations, clinical benefits, pacing thresholds, electrogram amplitudes and lead impedance..

Results: There were 20% lead re-implantations after two years of follow-up and a significant improvement in NYHA functional capacity. For the coronary venous lead the average stimulation threshold was stable at 1.2-1.3 V at 0.5 ms, the sensing electrogram amplitude was 10-18 mV and the impedance was 600 ohm at 5.0 V during follow-up. Similar stable electrophysiological values were measured for the right ventricular lead and the atrial lead.

Interpretation: There were in general few peroperative complications. All implanted leads in CRT showed stable- and low threshold values during a two-year follow-up period.

Download full-text PDF

Source

Publication Analysis

Top Keywords

resynchronization therapy
8
severe congestive
8
congestive heart
8
two-year follow-up
8
lead re-implantations
8
lead
5
[cardiac resynchronization
4
therapy severe
4
heart failure]
4
failure] background
4

Similar Publications

Background: Coronary sinus (CS) lead placement in persistent left superior vena cava (PLSVC) cases is challenging because of the poor backup force of the guiding catheter within the enlarged CS. Active fixation Quadripolar leads (Attain Stability™ Quad 4798, Medtronic) can expand choice to CS branches with limited access; however, no cases of anchoring to the main body of the CS have been published to date.

Case Summary: We describe a case of cardiac resynchronization therapy pacemaker upgrade in a 79-year-old female who developed pacing-induced cardiomyopathy after pacemaker implantation via the right superior vena cava (SVC) for atrioventricular block eight years ago wherein PLSVC was revealed during the procedure.

View Article and Find Full Text PDF

Circadian aspects in nonpharmacologic and pharmacologic treatment of insomnia.

Handb Clin Neurol

January 2025

Department of Surgical Medical and Molecular Pathology and Critical Care Medicine, University of Pisa, Pisa, Italy; Department of Neuroscience, Psychology Unit, University of Pisa Azienda Ospedaliera Universitaria Pisana (AUOP), Pisa, Italy.

Insomnia disorder is a frequent sleep disorder leading to significant health and economic consequences. It has been proposed that individuals with insomnia may experience compromised deactivation systems of arousal, leading to a chronic state of hyperactivation of arousal known as hyperarousal, along with instability in the flip-flop system. Such disruptions may have a primarily impact on the sleep homeostatic drive process.

View Article and Find Full Text PDF

How to correct QT interval after cardiac resynchronisation therapy.

J Electrocardiol

January 2025

Victorian Heart Institute, Monash University, Clayton, VIC, Australia; Victorian Heart Hospital, Clayton, VIC, Australia. Electronic address:

Introduction: This study evaluates various formulae used to correct the QT interval in patients with wide QRS complexes to calculate corrected QT (QTc) following Cardiac Resynchronisation Therapy (CRT).

Methods: We included patients with severe heart failure and left bundle branch block, presenting with a QRS duration of at least 120 milliseconds, who underwent successful CRT implantation. Patients were excluded if they had non-lateral left ventricular lead placement, metabolic disorders, atrial fibrillation, atrial tachycardia, or high-degree atrioventricular block prior to implantation.

View Article and Find Full Text PDF

Leadless Pacing: Current Status and Ongoing Developments.

Micromachines (Basel)

January 2025

Section of Electrophysiology, Division of Cardiology, Department of Internal Medicine, Rush University Medical Center, 1653 W. Congress, Chicago, IL 60612, USA.

Although significant strides have been made in cardiac pacing, the field is still evolving. While transvenous permanent pacing is highly effective in the management of bradyarrhythmias, it is not risk free and may result in significant morbidity and, rarely, mortality. Transvenous leads are often the weakest link in a pacing system.

View Article and Find Full Text PDF

Background: The benefit of implantable cardioverter-defibrillators (ICD) and cardiovascular resynchronization therapy (CRT-D) in patients supported with a HeartMate 3 left ventricular assist device (LVAD) remains uncertain.

Methods: An analysis of the MOMENTUM 3 randomized clinical trial and the first 1000 patients in the Continued Access Protocol trial. Patients were divided into three groups based on the presence of ICD and/or CRT-D: No device (n=153, 11%), ICD only (n=699, 50.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!